Pulse oximetry: fundamentals and technology update

Affiliations
Review

Pulse oximetry: fundamentals and technology update

Meir Nitzan et al. Med Devices (Auckl). .

Abstract

Oxygen saturation in the arterial blood (SaO2) provides information on the adequacy of respiratory function. SaO2 can be assessed noninvasively by pulse oximetry, which is based on photoplethysmographic pulses in two wavelengths, generally in the red and infrared regions. The calibration of the measured photoplethysmographic signals is performed empirically for each type of commercial pulse-oximeter sensor, utilizing in vitro measurement of SaO2 in extracted arterial blood by means of co-oximetry. Due to the discrepancy between the measurement of SaO2 by pulse oximetry and the invasive technique, the former is denoted as SpO2. Manufacturers of pulse oximeters generally claim an accuracy of 2%, evaluated by the standard deviation (SD) of the differences between SpO2 and SaO2, measured simultaneously in healthy subjects. However, an SD of 2% reflects an expected error of 4% (two SDs) or more in 5% of the examinations, which is in accordance with an error of 3%-4%, reported in clinical studies. This level of accuracy is sufficient for the detection of a significant decline in respiratory function in patients, and pulse oximetry has been accepted as a reliable technique for that purpose. The accuracy of SpO2 measurement is insufficient in several situations, such as critically ill patients receiving supplemental oxygen, and can be hazardous if it leads to elevated values of oxygen partial pressure in blood. In particular, preterm newborns are vulnerable to retinopathy of prematurity induced by high oxygen concentration in the blood. The low accuracy of SpO2 measurement in critically ill patients and newborns can be attributed to the empirical calibration process, which is performed on healthy volunteers. Other limitations of pulse oximetry include the presence of dyshemoglobins, which has been addressed by multiwavelength pulse oximetry, as well as low perfusion and motion artifacts that are partially rectified by sophisticated algorithms and also by reflection pulse oximetry.

Keywords: arterial blood; oxygen saturation; photoplethysmography; pulse oximetry; venous blood.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Absorption spectra of the oxygenated and deoxygenated hemoglobin molecules. Notes: In the red and the infrared regions, the absorption is relatively low and allows accurate measurement of light transmission. Copyright © 1999. Prahl S. Reproduced from Prahl S. Optical absorption of hemoglobin. 1999. Available from: http://omlc.ogi.edu/spectra/hemoglobin/index.html. Accessed May 26, 2014. Abbreviations: HbO2, oxygenated hemoglobin; Hb, deoxygenated hemoglobin.
Figure 2
Figure 2
The photoplethysmography signal. Note: DC denotes the pulse baseline and AC the pulse amplitude.

References

    1. Guyton AC, Hall JE. Textbook of Medical Physiology. 9th ed. Philadelphia: WB Saunders; 1996.
    1. Wagner PD. Ventilation, blood flow and gas exchange. In: Mason RJ, Broaddus VC, Murray JF, Nadel JA, editors. Murray and Nadel’s Textbook of Respiratory Medicine. Philadelphia: WB Saunders; 2005.
    1. Farmer J. Blood oxygen measurement. In: Webster JG, editor. Design of Pulse Oximeters. Bristol: Institute of Physics; 1997. pp. 21–39.
    1. Gold WM. Pulmonary function testing. In: Mason RJ, Broaddus VC, Murray JF, Nadel JA, editors. Murray and Nadel’s Textbook of Respiratory Medicine. Philadelphia: WB Saunders; 2005.
    1. Yoshiya I, Shimady Y, Tanake K. Spectrophotometric monitoring of arterial oxygen saturation on the fingertip. Med Biol Eng Comput. 1980;18(1):27–32. - PubMed