. 2014 Nov;21(11):1363-9.
doi: 10.1016/j.acra.2014.06.012. Epub 2014 Aug 12.

Use of contrast-enhanced spectral mammography for intramammary cancer staging: preliminary results

Affiliations

Use of contrast-enhanced spectral mammography for intramammary cancer staging: preliminary results

Katrin S Blum et al. Acad Radiol. 2014 Nov.

Abstract

Rationale and objectives: To prospectively evaluate and compare the accuracy of contrast-enhanced spectral mammography (CESM) and ultrasound (US) in size measurement of breast cancer with histologic tumor sizes as gold standard.

Materials and methods: Twenty women aged between 40-73 years (mean age, 57 ± 10 years) with histologically proven invasive ductal/lobular carcinomas were included in the study. Agreement between imaging tumor size (CESM and US) and histopathologic tumor size was evaluated with Bland-Altman analysis. Stereotactically guided vacuum biopsy was performed in four patients after CESM. Two independent reviewers described artifacts of CESM.

Results: Motion artifacts did not occur in the study. CESM-specific artifacts caused by scattered radiation mostly occurred in oblique view of CESM. Background enhancement of breast tissue was seen in four patients. Mean difference of tumor sizes was 0.3 mm (6.34%) between CESM and histology and -2.2 mm (-7.59%) between US and histology. Limits of agreement ranged from -18.9 to 19.48 mm for CESM and from -17.1 to 12.7 mm with US. Especially smaller tumors with a size <23 mm were measured more precisely with CESM. Enhancement of breast tissue around microcalcifications correlated with abnormalities.

Conclusions: CESM is accurate in size measurements of small breast tumors. On average CESM leads to a slight overestimation of tumor size, whereas US tends to underestimate tumor size. Assessment of the breast tissue can be limited by the scattered radiation artifact and background enhancement of breast tissue. CESM seems to be helpful in the characterization of breast tissue around microcalcifications.

Keywords: Contrast-enhanced spectral mammography; artifacts; tumor staging; ultrasound.