Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2014 Oct;7(10):1159-67.
doi: 10.1016/j.jcin.2014.04.016. Epub 2014 Aug 13.

Transapical transcatheter aortic valve for severe aortic regurgitation: expanding the limits

Affiliations
Free article

Transapical transcatheter aortic valve for severe aortic regurgitation: expanding the limits

Daniel Wendt et al. JACC Cardiovasc Interv. 2014 Oct.
Free article

Abstract

Objectives: This study sought to evaluate the self-expandable ACURATE TA device (Symetis SA, Ecublens, Switzerland) in a cohort of patients with pure aortic regurgitation (AR).

Background: Transcatheter aortic valve replacement (TAVR) has been initially considered as an alternative for high-risk patients with aortic stenosis. Although the current experience is limited, TAVR might be also an alternative to treat patients with pure, severe AR.

Methods: Between April 2012 and December 2013, a total of 8 high-risk patients with pure, severe AR were enrolled (grade III+). Clinical and hemodynamic data as well as data on device and procedure parameters and outcomes were collected.

Results: Patient mean was 72.5 ± 8.4 years, and 37.5% of patients were female. Logistic EuroSCORE was 34.0 ± 7.9% and the Society of Thoracic Surgeons score was 7.3 ± 3.3% on average. Two patients had undergone emergency aortic operation before due to acute type A aortic dissection, and both were treated by replacement of the ascending aorta (including root reconstruction) and the aortic arch combined with or without E-vita Open stent graft (Jotec GmbH, Hechingen, Germany) (January 2011 and March 2012), whereas the other patients experienced primary AR. All patients underwent successful transapical TAVR with the transapical ACURATE TA device (size small, n = 1, size medium, n = 3, size large, n = 4) without any intraprocedural complications according to the Valve Academic Research Consortium 2 criteria. Post-procedure AR grade I+ or lower, as revealed by transoesophageal echocardiography and angiography, was present in all 8 patients. At 30 days, the stroke incidence and all-cause mortality rate were 0%.

Conclusions: This small single-center series demonstrates the feasibility of transapical TAVR with the self-expandable ACURATE TA device in high-risk patients with severe AR.

Keywords: aortic regurgitation; aortic valve; high-risk; transcatheter aortic valve replacement.

PubMed Disclaimer

Comment in

MeSH terms

LinkOut - more resources