Observational Study
. 2014 Oct;40(10):1654-60.
doi: 10.1016/j.jcrs.2014.01.044. Epub 2014 Aug 20.

Long-term clinical outcomes of toric intraocular lens implantation in cataract cases with preexisting astigmatism

Affiliations
Observational Study

Long-term clinical outcomes of toric intraocular lens implantation in cataract cases with preexisting astigmatism

Toshiyuki Miyake et al. J Cataract Refract Surg. 2014 Oct.

Abstract

Purpose: To evaluate the long-term clinical outcomes and rotational stability of toric intraocular lenses (IOLs) to correct preexisting astigmatism in cataract patients.

Setting: Kitasato University Hospital, Kanagawa, Japan.

Design: Prospective observational study.

Methods: Phacoemulsification with implantation of a toric IOL (Acrysof IQ toric SN6AT) was performed in cataract patients with corneal astigmatism. The postoperative follow-up was up to 2 years.

Results: This study enrolled 378 eyes of 302 patients with a mean age of 63.4 years ± 16.9 (SD). The corrected distance visual acuity 3 months postoperatively was 20/25 or better in 94.7% of eyes. The mean refractive cylinder was -1.92 ± 1.45 diopters (D) preoperatively, -0.59 ± 0.62 D 3 months postoperatively (322 eyes), and -0.67 ± 0.90 D at 2 years (73 eyes). There were statistically significant differences between the preoperative and 3-month postoperative measurements (P<.001, Wilcoxon signed-rank test). The mean IOL misalignment was 4.1 ± 3.0 degrees 2 years postoperatively. The mean IOL rotation was 4.5 ± 4.9 degrees within 1 day postoperatively. The rotation was more than 20 degrees in 6 eyes, all of which had an axial length (AL) of more than 25.0 mm. All rotations occurred within 10 days postoperatively.

Conclusions: Toric IOLs were effective in reducing preexisting corneal astigmatism and had overall good rotational stability. A large degree of IOL rotation might occur in eyes with a relatively long AL, especially during the early postoperative period. The 6 rotated IOLs were implanted to correct with-the-rule astigmatism.

Financial disclosure: No author has a financial or proprietary interest in any material or method mentioned.

PubMed Disclaimer

Comment in

  • Preventing toric intraocular lens rotation.
    Yeoh R. Yeoh R. J Cataract Refract Surg. 2015 Jan;41(1):251-2. doi: 10.1016/j.jcrs.2014.11.006. J Cataract Refract Surg. 2015. PMID: 25532663 No abstract available.
  • Reply: To PMID 25149554.
    Miyake T, Kamiya K, Shimizu K. Miyake T, et al. J Cataract Refract Surg. 2015 Jan;41(1):252. doi: 10.1016/j.jcrs.2014.11.007. J Cataract Refract Surg. 2015. PMID: 25532665 No abstract available.
  • Toric intraocular lens rotation related to the capsulorhexis.
    Torquetti L. Torquetti L. J Cataract Refract Surg. 2015 Feb;41(2):483. doi: 10.1016/j.jcrs.2014.12.046. J Cataract Refract Surg. 2015. PMID: 25661158 No abstract available.
  • Reply: To PMID 25149554.
    Miyake T, Kamiya K, Shimizu K. Miyake T, et al. J Cataract Refract Surg. 2015 Feb;41(2):483. doi: 10.1016/j.jcrs.2014.12.045. J Cataract Refract Surg. 2015. PMID: 25661159 No abstract available.