Randomized Controlled Trial
. 2015 Apr;37(2):220-8.
doi: 10.1097/FTD.0000000000000127.

A validated method for quantification of efavirenz in dried blood spots using high-performance liquid chromatography-mass spectrometry

Affiliations
Randomized Controlled Trial

A validated method for quantification of efavirenz in dried blood spots using high-performance liquid chromatography-mass spectrometry

Alieu B Amara et al. Ther Drug Monit. 2015 Apr.

Abstract

Background: Efavirenz (EFV) is one of the preferred components of first-line antiretroviral treatment. EFV is characterized by a long plasma half-life (40-55 hours) with large interpatient variability, which raises the potential for individualization of therapy. Analyses of EFV levels in plasma require specialized facilities (cold storage/transport) which, in resource-limited settings, can be problematic; dried blood spots (DBS)-EFV measurements thus provide a cheap easy alternative for therapeutic drug monitoring. Our aim was to develop and validate a liquid chromatography-mass spectrometry method to quantify EFV in DBS collected as part of clinical trials in resource-limited settings.

Methods: DBS for standards, quality control samples, and patient samples were excised and then extracted with ethyl acetate/n-hexane (50/50 vol/vol) after addition of internal standard hexobarbital, and 1 mol/L K2CO3. The extract was evaporated to dryness, the residue reconstituted in mobile phase and analyzed directly by liquid chromatography-mass spectrometry. Gradient elution was on a reverse-phase C18 column using 1 mmol/L ammonium acetate in water and acetonitrile. Quantification was by selected reaction monitoring in negative ionization mode. DBS samples were obtained at several time points over 24 hours from HIV+ patients on either 400 or 600 mg EFV in combination with emtricitabine/tenofovir.

Results: The internal standard and EFV eluted at 2.68 and 3.54 minutes, respectively in a 5-minute run time. Matrix effects were minimal (-5.4%). Calibration curves were validated over a concentration range of 25-5000 ng/mL. Intra-assay and interassay variations ranged between 6.7% and 8.7% for imprecision and 100.3% and 104.2% for accuracy. Mean recovery was >64%. The DBS data showed a strong positive correlation with a validated plasma EFV assay (R = 0.9764, P < 0.001). EFV concentrations from DBS were approximately 42% lower than the paired plasma values, and the ratio of blood/plasma did not change over the dosing interval.

Conclusions: The validated assay is now routinely applied to clinical samples measuring DBS EFV for pharmacokinetic analysis. The methodology is robust, accurate, and sensitive.

PubMed Disclaimer

Publication types

LinkOut - more resources