. 2015 Jan;25(1):246-57.
doi: 10.1007/s00330-014-3391-7. Epub 2014 Sep 5.

Robotic-assisted thermal ablation of liver tumours

Affiliations

Robotic-assisted thermal ablation of liver tumours

Basri Johan Jeet Abdullah et al. Eur Radiol. 2015 Jan.

Abstract

Objective: This study aimed to assess the technical success, radiation dose, safety and performance level of liver thermal ablation using a computed tomography (CT)-guided robotic positioning system.

Methods: Radiofrequency and microwave ablation of liver tumours were performed on 20 patients (40 lesions) with the assistance of a CT-guided robotic positioning system. The accuracy of probe placement, number of readjustments and total radiation dose to each patient were recorded. The performance level was evaluated on a five-point scale (5-1: excellent-poor). The radiation doses were compared against 30 patients with 48 lesions (control) treated without robotic assistance.

Results: Thermal ablation was successfully completed in 20 patients with 40 lesions and confirmed on multiphasic contrast-enhanced CT. No procedure related complications were noted in this study. The average number of needle readjustment was 0.8 ± 0.8. The total CT dose (DLP) for the entire robotic assisted thermal ablation was 1382 ± 536 mGy.cm, while the CT fluoroscopic dose (DLP) per lesion was 352 ± 228 mGy.cm. There was no statistically significant (p > 0.05) dose reduction found between the robotic-assisted versus the conventional method.

Conclusion: This study revealed that robotic-assisted planning and needle placement appears to be safe, with high accuracy and a comparable radiation dose to patients.

Key points: • Clinical experience on liver thermal ablation using CT-guided robotic system is reported. • The technical success, radiation dose, safety and performance level were assessed. • Thermal ablations were successfully performed, with an average performance score of 4.4/5.0. • Robotic-assisted ablation can potentially increase capabilities of less skilled interventional radiologists. • Cost-effectiveness needs to be proven in further studies.

PubMed Disclaimer

References

    1. Int J Med Robot. 2014 Mar;10(1):35-43 - PubMed
    1. Acta Radiol. 1991 Jan;32(1):83-5 - PubMed
    1. Int J Hepatol. 2011;2011:104685 - PubMed
    1. Radiol Clin North Am. 2000 Mar;38(2):303-22, viii - PubMed