Accuracy of calculated arterial saturation in oxygen in neonatal foals and effects of monitor, sensor, site of sensor placement, and degree of hypoxemia on the accuracy of pulse oximetry
- PMID: 25212699
- DOI: 10.1111/vec.12231
Accuracy of calculated arterial saturation in oxygen in neonatal foals and effects of monitor, sensor, site of sensor placement, and degree of hypoxemia on the accuracy of pulse oximetry
Abstract
Objectives: To (1) assess the agreement between calculated values for arterial hemoglobin saturation of oxygen (SaO2 ) and values obtained by co-oximetry, (2) assess the accuracy of 2 pulse oximetry monitors for measurement of SaO2 in foals, and for each monitor, and (3) determine the optimal combination of sensor type and site of sensor placement for SaO2 monitoring.
Design: Prospective experimental study.
Setting: University teaching hospital.
Animals: Six neonatal foals.
Interventions: Foals were anesthetized with isoflurane and SaO2 was manipulated by varying the inspired fraction of oxygen. SaO2 was calculated from oxygen tension or measured by pulse oximetry using 2 monitors equipped with transmission or reflectance sensors attached to the foal's tongue, lip, ear, or inserted rectally (reflectance sensor only). SaO2 values measured by co-oximetry were used as the gold standard to calculate bias.
Measurements and main results: Mean (±SD) SaO2 determined by co-oximetry was 65.2% ± 11.8%, 85.4% ± 2.5%, and 97.2% ± 0.4% at the low, intermediate, and high SaO2 level, respectively. Sensors attached to the ear failed to provide SaO2 readings for most attempts. Reflectance sensors placed on the lip or rectally gave significantly larger biases at low SaO2 (-17.0% and -23.6%, respectively) than at higher levels. Bias of all other combinations of monitors, sensors, and sites (-1.8 to -4.1%) was not significantly influenced by the level of SaO2 or different from each other. The bias of calculated saturation was similar to that of pulse oximetry.
Conclusions: Transmission sensors placed on the lip or tongue or reflectance sensors placed on the tongue give the most accurate assessment of SaO2 in anesthetized neonatal foals. Calculated saturation is not more accurate than pulse oximetry to estimate SaO2 .
Keywords: anesthesia; blood gas analysis; oxygenation monitoring.
© Veterinary Emergency and Critical Care Society 2014.
Similar articles
-
Pulse oximetry at two sensor placement sites in conscious foals.Acta Vet Scand. 2025 Jan 23;67(1):6. doi: 10.1186/s13028-025-00794-w. Acta Vet Scand. 2025. PMID: 39849542 Free PMC article.
-
Evaluation of pulse oximetry in anaesthetised foals using multiple combinations of transducer type and transducer attachment site.Equine Vet J. 1996 Nov;28(6):437-45. doi: 10.1111/j.2042-3306.1996.tb01615.x. Equine Vet J. 1996. PMID: 9049492
-
Noninvasive oximetry in the rat.Biomed Instrum Technol. 1989 May-Jun;23(3):222-8. Biomed Instrum Technol. 1989. PMID: 2752232
-
The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis.BMC Med. 2022 Aug 16;20(1):267. doi: 10.1186/s12916-022-02452-8. BMC Med. 2022. PMID: 35971142 Free PMC article.
-
Advanced Uses of Pulse Oximetry for Monitoring Mechanically Ventilated Patients.Anesth Analg. 2017 Jan;124(1):62-71. doi: 10.1213/ANE.0000000000001283. Anesth Analg. 2017. PMID: 27183375 Review.
Cited by
-
Pulse oximetry at two sensor placement sites in conscious foals.Acta Vet Scand. 2025 Jan 23;67(1):6. doi: 10.1186/s13028-025-00794-w. Acta Vet Scand. 2025. PMID: 39849542 Free PMC article.
-
Reliability, clinical performance and trending ability of a pulse oximeter and pulse co-oximeter, in monitoring blood oxygenation, at two measurement sites, in immobilised white rhinoceros (Ceratotherium simum).BMC Vet Res. 2024 Jul 16;20(1):319. doi: 10.1186/s12917-024-04179-5. BMC Vet Res. 2024. PMID: 39014383 Free PMC article.
-
Bi-Level Positive Airway Pressure for Non-invasive Respiratory Support of Foals.Front Vet Sci. 2021 Sep 29;8:741720. doi: 10.3389/fvets.2021.741720. eCollection 2021. Front Vet Sci. 2021. PMID: 34660771 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources