Randomized Controlled Trial
. 2015 Dec;23(12):3637-45.
doi: 10.1007/s00167-014-3264-2. Epub 2014 Sep 13.

Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial

Affiliations
Randomized Controlled Trial

Comparison between patient-specific instruments and conventional instruments and computer navigation in total knee arthroplasty: a randomized controlled trial

Chun Hoi Yan et al. Knee Surg Sports Traumatol Arthrosc. 2015 Dec.

Abstract

Purpose: The current study investigated the accuracy in achieving proper lower limb alignment and individual component positions after total knee arthroplasty (TKA) with 3 different instrumentation techniques. It was hypothesized that patient-specific instruments (PSI) would achieve more accurate lower limb alignment and component positions compared to conventional instruments (CON).

Methods: Ninety knees in 81 patients were randomized in 1:1:1 ratio into CON, computer navigation (NAV) and PSI groups to receive TKA. The surgical routines were standardized. The lower limb mechanical axis and individual component positions were assessed on standard radiographs. Tourniquet time, operation time and patients' functional scores were documented.

Results: Conventional instruments and PSI were more likely to result in an excessively flexed femoral component (p = 0.001) compared to NAV. Number of outliers in postoperative lower limb alignment, and other components positions in the coronal and sagittal plane showed no statistically significant difference. The mean tourniquet time and operation time was significantly shorter in CON and PSI groups than NAV group (p < 0.001). Four early complications occurred in the PSI group (p = 0.015). At 3-month follow-up, there was no difference in terms of the knee range of motion and patients' function among the 3 groups.

Conclusion: No significant radiological and clinical benefit could be demonstrated in using PSI over CON or NAV in TKA. Routine use of PSI is not recommended because of the extra cost and waiting time.

Level of evidence: I.

Keywords: Component position; Computer navigation; Functional outcomes; Lower limb alignment; Patient-specific instrumentation; Total knee arthroplasty.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Lancet. 1986 Feb 8;1(8476):307-10 - PubMed
    1. Knee Surg Sports Traumatol Arthrosc. 2013 Oct;21(10):2194-9 - PubMed
    1. Orthopedics. 2008 Sep;31(9):927-30 - PubMed
    1. J Bone Joint Surg Br. 2004 Apr;86(3):372-7 - PubMed
    1. J Bone Joint Surg Br. 2004 Jul;86(5):682-7 - PubMed

Publication types