Comparison of double dose plateletpheresis on the Fenwal Amicus, Fresenius COM.TEC and Trima Accel cell separators
- PMID: 25219639
- DOI: 10.1016/j.transci.2014.08.022
Comparison of double dose plateletpheresis on the Fenwal Amicus, Fresenius COM.TEC and Trima Accel cell separators
Abstract
Background: A variety of apheresis instruments are now available on the market for double dose plateletpheresis. We compared three apheresis devices (Fenwal Amicus, Fresenius COM.TEC and Trima Accel) with regard to processing time, platelet (PLT) yield, collection efficiency (CE) and collection rate (CR).
Study design and methods: The single-needle or double-needle double plateletpheresis procedures of the three instruments were compared in a retrospective, randomized study in 135 donors.
Results: In the pre-apheresis setting, 45 double plateletpheresis procedures performed with each instrument revealed no significant differences in donor's age, sex, weight, hemoglobin, white blood cell and PLT count between three groups. The blood volume processed to reach a target PLT yield of ≥ 6 × 10(11) was higher in the COM.TEC compared with the Amicus and Trima (4394 vs. 3780 and 3340 ml, respectively; p < 0.001). Also there was a significantly higher median volume of ACD used in collections on the COM.TEC compared with the Amicus and Trima (426 vs. 387 and 329 ml, respectively; p < 0.001). There was a significantly higher median time needed for the procedures on the COM.TEC compared with the Amicus and Trima (66 vs. 62 and 63 min, respectively; p = 0.024). The CE was significantly higher with the Trima compared with the Amicus and COM.TEC (83.57 ± 17.19 vs. 66.71 ± 3.47 and 58.79 ± 5.14%, respectively; p < 0.001). Also, there was a significantly higher product volume on the Trima compared with the Amicus and COM.TEC (395.56 vs. 363.11 and 386.4 ml, respectively; p = 0.008). Additionally, the CR was significantly lower with the COM.TEC compared with the Amicus and Trima (0.092 ± 0.011 vs. 0.099 ± 0.013 and 0.097 ± 0.013 plt × 10(11)/min, respectively; p = 0.039). There was no significant differences in PLT yield between the three groups (p = 0.636).
Conclusions: Trima single-needle device collected double dose platelets more efficiently than Amicus and COM.TEC double-needle devices. Blood volume processed, ACD-A volume, and median separation time was significantly higher with the COM.TEC. Also, the CR was significantly lower with the COM.TEC.
Keywords: Amicus; Cell separator; Com.tec; Double plateletpheresis; Trima.
Copyright © 2014 Elsevier Ltd. All rights reserved.
Similar articles
-
Comparison of plateletpheresis on the Fenwal Amicus, Fresenius COM.TEC, and Trima Accel Cell separators.J Clin Apher. 2015 Jun;30(3):171-5. doi: 10.1002/jca.21358. Epub 2014 Sep 17. J Clin Apher. 2015. PMID: 25230340 Clinical Trial.
-
Comparison of Plateletpheresis on the Fenwal Amicus and Fresenius Com.Tec Cell Separators.Transfus Med Hemother. 2008;35(5):368-373. doi: 10.1159/000151351. Epub 2008 Sep 16. Transfus Med Hemother. 2008. PMID: 21512626 Free PMC article.
-
Effectiveness of the Trima Accel cell separator in the double dose plateletpheresis.Transfus Apher Sci. 2016 Oct;55(2):240-242. doi: 10.1016/j.transci.2016.07.022. Epub 2016 Aug 2. Transfus Apher Sci. 2016. PMID: 27522105 Clinical Trial.
-
Blood component collection by apheresis.J Clin Apher. 2006 Jul;21(2):142-51. doi: 10.1002/jca.20043. J Clin Apher. 2006. PMID: 15880369 Review.
-
Plateletpheresis: what's new?Transfus Sci. 1996 Dec;17(4):537-44. Transfus Sci. 1996. PMID: 10168551 Review.
Cited by
-
Analysis of Donor Safety in High Yield Plateletpheresis Procedures: An Experience from Tertiary Care Hospital in South India.Indian J Hematol Blood Transfus. 2020 Jul;36(3):542-549. doi: 10.1007/s12288-020-01266-z. Epub 2020 Feb 27. Indian J Hematol Blood Transfus. 2020. PMID: 32647430 Free PMC article.
-
Double Dose Plateletpheresis: A Savior to Shrinking Donor Pool and Platelet Inventory Management.Indian J Hematol Blood Transfus. 2018 Oct;34(4):691-696. doi: 10.1007/s12288-018-0920-6. Epub 2018 Jan 20. Indian J Hematol Blood Transfus. 2018. PMID: 30369742 Free PMC article.
-
Effect of double dose plateletpheresis on target yield and donor platelet recovery.Hematol Transfus Cell Ther. 2023 Jan-Mar;45(1):16-24. doi: 10.1016/j.htct.2021.05.002. Epub 2021 Jun 18. Hematol Transfus Cell Ther. 2023. PMID: 34294601 Free PMC article.
-
Feasibility of high-yield plateletpheresis in routine practice: Experience from tertiary health center from South India.Asian J Transfus Sci. 2023 Jan-Jun;17(1):34-40. doi: 10.4103/ajts.ajts_103_21. Epub 2022 Sep 28. Asian J Transfus Sci. 2023. PMID: 37188018 Free PMC article.
-
Plastic-based acoustofluidic devices for high-throughput, biocompatible platelet separation.Lab Chip. 2019 Jan 29;19(3):394-402. doi: 10.1039/c8lc00527c. Lab Chip. 2019. PMID: 30631874 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources