Clinical evaluation of a high-fidelity wireless intravaginal pressure sensor
- PMID: 25224148
- DOI: 10.1007/s00192-014-2500-0
Clinical evaluation of a high-fidelity wireless intravaginal pressure sensor
Abstract
Introduction and hypothesis: A wireless intravaginal pressure sensor (IVPS) has been developed to quantify abdominal pressure (P(abd)) changes during exercise and activities of daily living to guide post-operative advice given to women. In this pilot study, we aimed to compare IVPS performance, comfort, retention, and acceptability to a standard fluid-filled intrarectal pressure catheter currently used to measure P(abd) during routine urodynamics.
Methods: A Life-Tech 3-mm urodynamic intrarectal catheter and IVPS were inserted concurrently in volunteers attending a urodynamics clinic. The IVPS was positioned above the levator plate and the intrarectal catheter positioned in routine fashion well above the anal sphincter. Routine urodynamics was undertaken, with women asked to perform star jumps if coughing or Valsalva did not invoke leakage. Subjects rated device comfort using a visual analogue scale (0-10). Repeated measures Bland-Altman analysis determined level of agreement (LOA) between the two devices for peak pressures for each activity.
Results: Twenty-seven women were recruited, 67% of the participants preferred the IVPS, 18% the intrarectal catheter, while 15% had no preference. Mean comfort score was 0.9 ± 1.7 and 2.1 ± 2.6 (p = 0.049) for the IVPS and intrarectal catheter respectively. Bland-Altman analysis demonstrated minimal bias for cough and Valsalva, however LOA were wide. Differences were more prominent during star jumps where rapid dynamic pressure changes occurred.
Conclusions: The IVPS had a higher comfort score and was well retained. The LOA between the two systems was moderate, but the high sampling rate and lower susceptibility to motion artefacts of the IVPS may provide more accurate information that will be important clinically.
Similar articles
-
Development and testing of a vaginal pressure sensor to measure intra-abdominal pressure in women.Neurourol Urodyn. 2010 Apr;29(4):532-5. doi: 10.1002/nau.20794. Neurourol Urodyn. 2010. PMID: 19693948 Free PMC article.
-
A critical appraisal of the methods of measuring leak-point pressures in women with stress incontinence.Obstet Gynecol. 1995 Sep;86(3):349-52. doi: 10.1016/0029-7844(95)00173-O. Obstet Gynecol. 1995. PMID: 7651641 Clinical Trial.
-
Urodynamic appraisal of vaginal versus rectal pressure recordings as indication of intra-abdominal pressure changes.Urology. 1986 May;27(5):482-5. doi: 10.1016/0090-4295(86)90424-3. Urology. 1986. PMID: 3705289
-
Ambulatory urodynamic monitoring: state of the art and future directions.Nat Rev Urol. 2019 May;16(5):291-301. doi: 10.1038/s41585-019-0175-5. Nat Rev Urol. 2019. PMID: 30936468 Free PMC article. Review.
-
Advances in Ambulatory Urodynamics.Curr Urol Rep. 2020 Aug 18;21(10):41. doi: 10.1007/s11934-020-00989-w. Curr Urol Rep. 2020. PMID: 32809183 Review.
Cited by
-
The Relationship Between Intra-Abdominal Pressure and Body Acceleration During Exercise.Female Pelvic Med Reconstr Surg. 2019 May/Jun;25(3):231-237. doi: 10.1097/SPV.0000000000000523. Female Pelvic Med Reconstr Surg. 2019. PMID: 29135811 Free PMC article.
-
Is Physical Activity Good or Bad for the Female Pelvic Floor? A Narrative Review.Sports Med. 2020 Mar;50(3):471-484. doi: 10.1007/s40279-019-01243-1. Sports Med. 2020. PMID: 31820378 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources