Validation of the Sysmex CS5100 coagulation analyzer and comparison to the Stago STA-R analyzer for routine coagulation parameters

Affiliations

Validation of the Sysmex CS5100 coagulation analyzer and comparison to the Stago STA-R analyzer for routine coagulation parameters

T Geens et al. Int J Lab Hematol. 2015 Jun.

Abstract

Introduction: The CS5100 analyzer (Sysmex) was validated for the determination of routine coagulation parameters. This fully automated coagulation analyzer uses multiple wavelength technology to perform coagulation (e.g., activated partial thromboplastin time - APTT, prothrombin time - PT, fibrinogen - FBG), chromogenic (e.g., antithrombin - AT) and immunological (e.g., D-dimers - DDi) assays.

Methods: A comparison with the currently used STA-R Evolution (Stago) was performed. Validation and verification of reference values of the CS5100 was performed in accordance to CLSI guidelines (H57-A, H47-A2, and C28-A3).

Results: As a different detection system and reagents were used, significant differences were observed (e.g. APTT). The within-day and between-day imprecision, accuracy and total error were all acceptable. The reference values defined by the manufacturer could be used except for APTT. In our settings, the therapeutic anti-Xa range of 0.3-0.7 IU/mL corresponded to an APTT range of 60-100 s (Dade actin FS reagent). The APTT reagent showed factor sensitivities between 46 and 72% for FVIII, IX, XI and XII while the PT reagent showed sensitivities between 34 and 52% for FII, FV, FXII, and FX.

Conclusion: In conclusion, the CS5100 instrument is suitable for the determination of the APTT, PT, FBG, DDi and AT in routine analysis.

Keywords: Stago STA-R; Sysmex CS5100; factor sensitivity; reference interval; unfractionated heparin range.