Intraoperative round window electrocochleography and speech perception outcomes in pediatric cochlear implant recipients
- PMID: 25259669
- DOI: 10.1097/AUD.0000000000000106
Intraoperative round window electrocochleography and speech perception outcomes in pediatric cochlear implant recipients
Abstract
Objectives: The goal was to measure the magnitude of cochlear responses to sound in pediatric cochlear implant recipients at the time of implantation and to correlate this magnitude with subsequent speech perception outcomes.
Design: A longitudinal cohort study of pediatric cochlear implant recipients was undertaken. Intraoperative electrocochleographic (ECoG) recordings were obtained from the round window in response to a frequency series at 90 dB nHL in 77 children totaling 89 ears (12 were second side surgeries) just before device insertion. The increase in intraoperative time was approximately 10 min. An ECoG "total response" metric was derived from the summed magnitudes of significant responses to the first, second, and third harmonics across a series of frequencies. A subset of these children reached at least 9 months of implant use and were old enough for the phonetically balanced kindergarten (PB-k) word test to be administered (n = 26 subjects and 28 ears). PB-k scores were compared to the ECoG total response and other biologic and audiologic variables using univariate and multiple linear regression analyses.
Results: ECoG responses were measurable in almost all ears (87 of 89). The range of ECoG total response covered about 60 dB (from ~0.05 to 50 μV). Analyzing individual ECoG recordings in bilaterally implanted children revealed poor concordance between the measured response in the first versus second ear implanted (r = 0.21; p = 0.13; n = 12). In a univariate linear regression, the ECoG total response was significantly correlated with PB-k scores in the subset of 26 subjects who were able to be tested and accounted for 32% of the variance (p = 0.002, n = 28). Preoperative pure-tone average (PTA) accounted for slightly more of the variance (r = 0.37, p = 0.001). However, ECoG total response and PTA were significantly but only weakly correlated (r = 0.14, p = 0.001). Other significant predictors of speech performance included hearing stability (stable versus progressive) and age at testing (22 and 16% of the variance, respectively). In multivariate analyses with these four factors, the ECoG accounted for the most weight (β = 0.36), followed by PTA (β = 0.26). In a hierarchical multiple regression analysis, the most parsimonious models that best predicted speech perception outcomes included three variables: ECoG total response, and any two of preoperative PTA, age at testing, or hearing stability. The various three factor models each predicted approximately 50% of the variance in word scores. Without the ECoG total response, the other three factors predicted 36% of variance.
Conclusions: Intraoperative round window ECoG recordings are reliably and easily obtained in pediatric cochlear implant recipients. The ECoG total response is significantly correlated with speech perception outcomes in pediatric implant recipients and can account for a comparable or greater proportion of variance in speech perception than other bio-audiologic factors. Intraoperative recordings can potentially provide useful prognostic information about acquisition of open set speech perception in implanted children.
Similar articles
-
Round window electrocochleography and speech perception outcomes in adult cochlear implant subjects: comparison with audiometric and biographical information.Otol Neurotol. 2014 Oct;35(9):e245-52. doi: 10.1097/MAO.0000000000000557. Otol Neurotol. 2014. PMID: 25118584
-
Round window electrocochleography just before cochlear implantation: relationship to word recognition outcomes in adults.Otol Neurotol. 2014 Jan;35(1):64-71. doi: 10.1097/MAO.0000000000000219. Otol Neurotol. 2014. PMID: 24317211 Free PMC article.
-
Round window electrocochleography before and after cochlear implant electrode insertion.Laryngoscope. 2016 May;126(5):1193-200. doi: 10.1002/lary.25602. Epub 2015 Sep 11. Laryngoscope. 2016. PMID: 26360623 Free PMC article.
-
Cochlear implantation in children with anomalous cochleovestibular anatomy.Laryngoscope. 2005 Jan;115(1 Pt 2 Suppl 106):1-26. doi: 10.1097/00005537-200501001-00001. Laryngoscope. 2005. PMID: 15626926 Review.
-
[Cochlear implant in children: rational, indications and cost/efficacy].Minerva Pediatr. 2013 Jun;65(3):325-39. Minerva Pediatr. 2013. PMID: 23685383 Review. Italian.
Cited by
-
Clinical Applications of Intracochlear Electrocochleography in Cochlear Implant Users With Residual Acoustic Hearing.J Audiol Otol. 2024 Apr;28(2):100-106. doi: 10.7874/jao.2024.00129. Epub 2024 Apr 10. J Audiol Otol. 2024. PMID: 38695055 Free PMC article.
-
On the Origin of the 1,000 Hz Peak in the Spectrum of the Human Tympanic Electrical Noise.Front Neurosci. 2017 Jul 11;11:395. doi: 10.3389/fnins.2017.00395. eCollection 2017. Front Neurosci. 2017. PMID: 28744193 Free PMC article.
-
Response Changes During Insertion of a Cochlear Implant Using Extracochlear Electrocochleography.Ear Hear. 2018 Nov/Dec;39(6):1146-1156. doi: 10.1097/AUD.0000000000000571. Ear Hear. 2018. PMID: 29554036 Free PMC article.
-
Intracochlear Electrocochleography and Speech Perception Scores in Cochlear Implant Recipients.Laryngoscope. 2021 Oct;131(10):E2681-E2688. doi: 10.1002/lary.29629. Epub 2021 May 21. Laryngoscope. 2021. PMID: 34019310 Free PMC article.
-
Residual Hair Cell Responses in Electric-Acoustic Stimulation Cochlear Implant Users with Complete Loss of Acoustic Hearing After Implantation.J Assoc Res Otolaryngol. 2021 Apr;22(2):161-176. doi: 10.1007/s10162-021-00785-4. Epub 2021 Feb 4. J Assoc Res Otolaryngol. 2021. PMID: 33538936 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous