Abstract
Background: The use of pediatric CT that had recently emerged as a valuable imaging tool has increased rapidly with an annual growth estimated at about 10% per year. Worldwide, there is a remarkable increase in the number of CT examinations performed. The purposes of this study are to: (i) to measure the radiation dose and estimate the effective doses to pediatric patients during CT for chest, abdomen and brain.
Material/methods: A total of 182 patients were investigated. CT scanners that participated in this study are helical CT scanners (64 slices, 16 slices and dual slices). Organ and surface dose to specific radiosensitive organs were estimated by using software from National Radiological Protection Board (NRPB).
Results: For all patients, the age was ranged between 1.12 month-10.0 years while the weight was ranged between 5.0 kg to 29.0 kg. The DLP was 320.58 mGy·cm, 79.93 mGy·cm, 66.63 mGy·cm for brain, abdomen and chest respectively. The effective dose was, 2.05, 1.8, 1.08 mSv for brain, abdomen and chest respectively.
Conclusions: The patient dose is independent of CT modality and depends on operator experience and CT protocol. The study has shown a great need for referring criteria, continuous training of staff in radiation protection concepts. Further studies are required in order to establish a reference level in Sudan.
Keywords: Dose-Response Relationship; Radiation; Relative Biological Effectiveness; Tomography Scanners; X-Ray Computed.
Similar articles
-
Establishment of institutional diagnostic reference level for computed tomography with automated dose-tracking software.J Med Radiat Sci. 2017 Jun;64(2):82-89. doi: 10.1002/jmrs.210. Epub 2017 Mar 1. J Med Radiat Sci. 2017. PMID: 28247587 Free PMC article.
-
Radiation doses to patients receiving computed tomography examinations in British Columbia.Can Assoc Radiol J. 2006 Apr;57(2):79-85. Can Assoc Radiol J. 2006. PMID: 16944681
-
Organ doses for reference pediatric and adolescent patients undergoing computed tomography estimated by Monte Carlo simulation.Med Phys. 2012 Apr;39(4):2129-46. doi: 10.1118/1.3693052. Med Phys. 2012. PMID: 22482634 Free PMC article.
-
Evaluation of radiation doses delivered in different chest CT protocols.Pol J Radiol. 2014 Jan;79:1-5. doi: 10.12659/PJR.889952. Pol J Radiol. 2014. PMID: 24454417 Free PMC article.
-
Radiation Doses in Consecutive CT Examinations from Five University of California Medical Centers.Radiology. 2015 Oct;277(1):134-41. doi: 10.1148/radiol.2015142728. Epub 2015 May 19. Radiology. 2015. PMID: 25988262 Free PMC article.
Cited by
-
Use of computed tomography and diffusion weighted imaging in children with ventricular shunt.Childs Nerv Syst. 2019 Mar;35(3):477-486. doi: 10.1007/s00381-018-04046-3. Epub 2019 Jan 8. Childs Nerv Syst. 2019. PMID: 30617617
References
-
- Seeram E. Computed Tomography: Physical Principles, Clinical Applications, and Quality Control. 3rd edition. W.B. Saunders Co; USA: 2008.
-
- International Atomic Energy Agency. Radiology Protection of Patients in Diagnostic and Interventional Radiology, Nuclear Medicine and Radiotherapy. Proceedings of an International Conference; Malaga, Spain. 2001.
-
- Amis ES, Butler PF, Applegate KE, et al. American College of Radiology white paper on radiation dose in medicine. J Am Coll Radiol. 2007;4(5):272–84. - PubMed
-
- Arch ME, Frush DP. Pediatric body MDCT: A 5-year follow-up survey of scanning parameters used by pediatric radiologists. Am J Roentgenol. 2008;191(2):611–17. - PubMed
-
- International Commission on Radiological Protection. Ann ICRP. 2–4. Vol. 37. Pergamon; Oxford, UK: 2007. The 2007 Recommendations of the International Commission on Radiological Protection. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources