Wear of highly crosslinked polyethylene acetabular components
- PMID: 25301435
- PMCID: PMC4404765
- DOI: 10.3109/17453674.2014.972890
Wear of highly crosslinked polyethylene acetabular components
Abstract
Background and purpose: Wear rates of highly crosslinked polyethylene (XLPE) acetabular components have varied considerably between different published studies. This variation is in part due to the different techniques used to measure wear and to the errors inherent in measuring the relatively low amounts of wear in XLPE bearings. We undertook a scoping review of studies that have examined the in vivo wear of XLPE acetabular components using the most sensitive method available, radiostereometric analysis (RSA).
Methods: A systematic search of the PubMed, Scopus, and Cochrane databases was performed to identify published studies in which RSA was used to measure wear of XLPE components in primary total hip arthroplasty (THA).
Results: 18 publications examined 12 primary THA cohorts, comprising only 260 THAs at 2-10 years of follow-up. The mean or median proximal wear rate reported ranged from 0.00 to 0.06 mm/year. However, differences in the manner in which wear was determined made it difficult to compare some studies. Furthermore, differences in RSA methodology between studies, such as the use of supine or standing radiographs and the use of beaded or unbeaded reference segments, may limit future meta-analyses examining the effect of patient and implant variables on wear rates.
Interpretation: This scoping review confirmed the low wear rates of XLPE in THA, as measured by RSA. We make recommendations to enhance the standardization of reporting of RSA wear results, which will facilitate early identification of poorly performing implants and enable a better understanding of the effects of surgical and patient factors on wear.
Figures
References
-
- AOANJRR. Annual Report. Adelaide: AOA; 2013. Australian Orthopaedic Association National Joint Replacement Registry.
-
- Ayers DC, Hays PL, Drew JM, Eskander MS, Osuch D, Bragdon CR. Two-year radiostereometric analysis evaluation of femoral head penetrati... . J Arthroplasty. 2009;24(6 Suppl):9–14. - PubMed
-
- Baldursson H, Egund N, Hansson LI, Selvik G. Instability and wear of total hip prostheses determined with roentgen st... . Arch Orthop Trauma Surg. 1979;95(4):257–63. - PubMed
-
- Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene. ... . J Bone Joint Surg Am. 2008;90(7):1487–91. - PubMed
-
- Borlin N, Rohrl SM, Bragdon CR. RSA wear measurements with or without markers in total hip arthroplasty . J Biomech. 2006;39(9):1641–50. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical