Comparative Study
. 2015 Apr;291(4):889-95.
doi: 10.1007/s00404-014-3509-9. Epub 2014 Oct 14.

Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography

Affiliations
Comparative Study

Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography

Michael Golatta et al. Arch Gynecol Obstet. 2015 Apr.

Abstract

Purpose: Automated three-dimensional (3D) breast ultrasound (US) systems are meant to overcome the shortcomings of hand-held ultrasound (HHUS). The aim of this study is to analyze and compare clinical performance of an automated 3D-US system by comparing it with HHUS, mammography and the clinical gold standard (defined as the combination of HHUS, mammography and-if indicated-histology).

Methods: Nine hundred and eighty three patients (=1,966 breasts) were enrolled in this monocentric, explorative and prospective cohort study. All examinations were analyzed blinded to the patients´ history and to the results of the routine imaging. The agreement of automated 3D-US with HHUS, mammography and the gold standard was assessed with kappa statistics. Sensitivity, specificity and positive and negative predictive value were calculated to assess the test performance.

Results: Blinded to the results of the gold standard the agreement between automated 3D-US and HHUS or mammography was fair, given by a Kappa coefficient of 0.31 (95% CI [0.26;0.36], p < 0.0001) and 0.25 (95% CI [0.2;0.3], p < 0.0001), respectively. Our results showed a high negative predictive value (NPV) of 98%, a high specificity of 85% and a sensitivity of 74% based on the cases with US-guided biopsy. Including the cases where the lesion was seen in a second-look automated 3D-US the sensitivity improved to 84% (NPV = 99%, specificity = 85%).

Conclusion: The results of this study let us suggest, that automated 3D-US might be a helpful new tool in breast imaging, especially in screening.