Parallax-free exophthalmometry: a comprehensive review of the literature on clinical exophthalmometry and the introduction of the first parallax-free exophthalmometer
- PMID: 25313439
- DOI: 10.3109/01676830.2014.963877
Parallax-free exophthalmometry: a comprehensive review of the literature on clinical exophthalmometry and the introduction of the first parallax-free exophthalmometer
Abstract
Purpose: To present the first parallax-free exophthalmometer design.
Background: Exophthalmometry is an important clinical tool. We provide a historic overview of clinical exophthalmometer designs, and we review current problems encountered in exophthalmometry.
Methods: We present a new and parallax-free exophthalmometer design that we have evaluated in 49 patients visiting our orbital clinic.
Results: The mean age of the patients was 49.8 years and 72% were female. The Pearson interobserver variation was 0.97, and 94% of the Hertel values measured by the two observers were within the limits (1.6 mm) of agreement.
Conclusion: This meter appears to be a reliable instrument for exophthalmometry. It is the first instrument that allows for a complete parallax-free measurement.
Keywords: Exophthalmometer; exophthalmometry; exophthalmos; parallax error; proptosis.
Similar articles
-
Reliability of exophthalmos measurement and the exophthalmometry value distribution in a healthy Dutch population and in Graves' patients. An exploratory study.Orbit. 2004 Sep;23(3):161-8. doi: 10.1080/01676830490504089. Orbit. 2004. PMID: 15545129
-
Proptosis--Correlation and Agreement between Hertel Exophthalmometry and Computed Tomography.Orbit. 2015;34(5):257-62. doi: 10.3109/01676830.2015.1057291. Epub 2015 Jul 17. Orbit. 2015. PMID: 26186249
-
Intra-observer and inter-observer variation of Hertel exophthalmometry.Ophthalmic Physiol Opt. 2009 Jul;29(4):472-6. doi: 10.1111/j.1475-1313.2008.00617.x. Ophthalmic Physiol Opt. 2009. PMID: 19523092 Clinical Trial.
-
Errors of single-mirror or prism Hertel exophthalmometers and recommendations for minimizing the errors.Ophthalmic Plast Reconstr Surg. 2007 May-Jun;23(3):197-201. doi: 10.1097/IOP.0b013e3180500d70. Ophthalmic Plast Reconstr Surg. 2007. PMID: 17519656
-
Normal Exophthalmometry Values in Iranian Population: A Meta-analysis.J Ophthalmic Vis Res. 2021 Jul 29;16(3):470-477. doi: 10.18502/jovr.v16i3.9441. eCollection 2021 Jul-Sep. J Ophthalmic Vis Res. 2021. PMID: 34394874 Free PMC article. Review.
Cited by
-
Association of anthropometric markers with globe position: A population-based MRI study.PLoS One. 2019 Feb 7;14(2):e0211817. doi: 10.1371/journal.pone.0211817. eCollection 2019. PLoS One. 2019. PMID: 30730926 Free PMC article.
-
Three-dimensional characterisation of the globe position in the orbit.Graefes Arch Clin Exp Ophthalmol. 2020 Jul;258(7):1527-1532. doi: 10.1007/s00417-020-04631-w. Epub 2020 Mar 5. Graefes Arch Clin Exp Ophthalmol. 2020. PMID: 32140924
-
Enophthalmos: Historical Perspective on Definitions, Measurement Devices, and Clinical Significance.Craniomaxillofac Trauma Reconstr. 2024 Dec;17(4):324-330. doi: 10.1177/19433875241236343. Epub 2024 Feb 23. Craniomaxillofac Trauma Reconstr. 2024. PMID: 39553789 Free PMC article.
-
Evaluation of Globe Position Asymmetry in Endocrine Orbitopathy Patients and a Control Group - A Computed Tomography-Based 3D Cephalometric Analysis.Eplasty. 2023 Feb 16;23:e19. eCollection 2023. Eplasty. 2023. PMID: 37187872 Free PMC article.
-
Visual outcomes endorse surgery of patients with spheno-orbital meningioma with minimal visual impairment or hyperostosis.Acta Neurochir (Wien). 2021 Jan;163(1):73-82. doi: 10.1007/s00701-020-04554-9. Epub 2020 Sep 4. Acta Neurochir (Wien). 2021. PMID: 32888075 Free PMC article.