Comparative Study
. 2015 Mar;25(3):687-93.
doi: 10.1007/s00330-014-3459-4. Epub 2014 Oct 15.

Assessment of image quality and low-contrast detectability in abdominal CT of obese patients: comparison of a novel integrated circuit with a conventional discrete circuit detector at different tube voltages

Affiliations
Comparative Study

Assessment of image quality and low-contrast detectability in abdominal CT of obese patients: comparison of a novel integrated circuit with a conventional discrete circuit detector at different tube voltages

A Euler et al. Eur Radiol. 2015 Mar.

Abstract

Objectives: To compare image quality and low-contrast detectability of an integrated circuit (IC) detector in abdominal CT of obese patients with conventional detector technology at low tube voltages.

Methods: A liver phantom with 45 lesions was placed in a water container to mimic an obese patient and examined on two different CT systems at 80, 100 and 120 kVp. The systems were equipped with either the IC or conventional detector. Image noise was measured, and the contrast-to-noise-ratio (CNR) was calculated. Low-contrast detectability was assessed independently by three radiologists. Radiation dose was estimated by the volume CT dose index (CTDIvol).

Results: The image noise was significantly lower, and the CNR was significantly higher with the IC detector at 80, 100 and 120 kVp, respectively (P = 0.023). The IC detector resulted in an increased lesion detection rate at 80 kVp (38.1 % vs. 17.2 %) and 100 kVp (57.0 % vs. 41.0 %). There was no difference in the detection rate between the IC detector at 100 kVp and the conventional detector at 120 kVp (57.0 % vs. 62.2 %). The CTDIvol at 80, 100 and 120 kVp measured 4.5-5.2, 7.3-7.9 and 9.8-10.2 mGy, respectively.

Conclusions: The IC detector at 100 kVp resulted in similar low-contrast detectability compared to the conventional detector with a 120-kVp protocol at a radiation dose reduction of 37 %.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Acad Radiol. 2006 Oct;13(10):1187-93 - PubMed
    1. Gastroenterol Clin North Am. 2010 Mar;39(1):1-7 - PubMed
    1. Radiology. 2009 Nov;253(2):520-31 - PubMed
    1. J Urol. 2013 Jun;189(6):2142-6 - PubMed
    1. Radiology. 2005 Aug;236(2):565-71 - PubMed

Publication types

LinkOut - more resources