Treating the patients in the 'grey-zone' with aortic valve disease: a comparison among conventional surgery, sutureless valves and transcatheter aortic valve replacement
- PMID: 25320140
- DOI: 10.1093/icvts/ivu340
Treating the patients in the 'grey-zone' with aortic valve disease: a comparison among conventional surgery, sutureless valves and transcatheter aortic valve replacement
Abstract
Objectives: Although the use of transcatheter aortic valve replacement (TAVR) has recently become an attractive strategy in extremely high-risk patients undergoing aortic valve replacement (AVR), the most appropriate treatment option in patients with an intermediate- to high-risk profile with conventional surgery (sAVR), TAVR or novel options, such as sutureless valves, has been widely debated.
Methods: One hundred and sixty-three consecutive patients with intermediate to high risk were prospectively enrolled and selected to undergo sAVR (Group 1: G1, n = 55), sutureless valve implantation (Group 2: G2, n = 53) or TAVR (Group 3: G3, n = 55) following a multidisciplinary evaluation including frailty, anatomy and degree of atherosclerotic disease of the aorta/peripheral vessels. The mean logistic EuroSCORE (G1 = 21.3 ± 12.7 vs G2 = 16 ± 11.7 vs G3 = 20.4 ± 12.7, P = 0.06) and preoperative demographics, such as age, gender and left ventricular ejection fraction, were similar: of note, chronic obstructive pulmonary disease was more frequent in TAVI patients (G1 = 27.2% vs G2 = 15.1% vs G3 = 47%; P <0.01). The Perceval S sutureless valve was used in Group 2, whereas TAVR was performed with a Corevalve prosthesis.
Results: Post-procedural pacemaker implantation (G1 = 1.8% vs G2 = 2% vs G3 = 25.5%, P <0.001) and peripheral vascular complications (G1 = 0% vs G2 = 0% vs G3 = 14.5%, P <0.001) occurred more frequently in patients undergoing TAVR. Hospital mortality was similar among the groups (G1 = 0% vs G2 = 0% vs G3 = 1.8%, P = NS). At the 24-month follow-up, overall survival free from major adverse cardiac and cerebrovascular events and prosthetic regurgitation was better in patients who had undergone sAVR and sutureless valves than those who had undergone TAVR (G1 = 95.2 ± 3.3% vs G2 = 91.6 ± 3.8% vs G3 = 70.5 ± 7.6%; P = 0.015).
Conclusions: This preliminary study suggests that the use of TAVR in patients with an intermediate- to high-risk profile is associated with a higher rate of perioperative complications and decreased survival at the 24-month follow-up compared with the use of conventional surgery or sutureless valves.
Keywords: Aortic valve replacement; Complications; Heart valve; Percutaneous; Surgery.
© The Author 2014. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
Similar articles
-
A Comparative Study of TAVR versus SAVR in Moderate and High-Risk Surgical Patients: Hospital Outcome and Midterm Results.Heart Surg Forum. 2019 Aug 27;22(5):E331-E339. doi: 10.1532/hsf.2243. Heart Surg Forum. 2019. PMID: 31596707
-
Conventional surgery, sutureless valves, and transapical aortic valve replacement: what is the best option for patients with aortic valve stenosis? A multicenter, propensity-matched analysis.J Thorac Cardiovasc Surg. 2013 Nov;146(5):1065-70; discussion 1070-1. doi: 10.1016/j.jtcvs.2013.06.047. Epub 2013 Sep 8. J Thorac Cardiovasc Surg. 2013. PMID: 24021955
-
Surgical sutureless and transcatheter aortic valves: hemodynamic performance and clinical outcomes in propensity score-matched high-risk populations with severe aortic stenosis.JACC Cardiovasc Interv. 2015 Apr 27;8(5):670-7. doi: 10.1016/j.jcin.2014.10.029. JACC Cardiovasc Interv. 2015. PMID: 25946438
-
Meta-analysis of complications in aortic valve replacement: comparison of Medtronic-Corevalve, Edwards-Sapien and surgical aortic valve replacement in 8,536 patients.Catheter Cardiovasc Interv. 2012 Jul 1;80(1):128-38. doi: 10.1002/ccd.23368. Epub 2012 Mar 13. Catheter Cardiovasc Interv. 2012. PMID: 22415849 Review.
-
What is the best approach in a patient with a failed aortic bioprosthetic valve: transcatheter aortic valve replacement or redo aortic valve replacement?Interact Cardiovasc Thorac Surg. 2015 Jun;20(6):837-43. doi: 10.1093/icvts/ivv037. Epub 2015 Mar 8. Interact Cardiovasc Thorac Surg. 2015. PMID: 25754372 Review.
Cited by
-
Sutureless Aortic Valve Replacement with Perceval Bioprosthesis Superior to Transcatheter Aortic Valve Implantation: A Promising Option for the Gray-Zone of Aortic Valve Replacement Procedures-A State-of-the-Art Systematic Review, Meta-Analysis, and Future Directions.J Clin Med. 2024 Aug 19;13(16):4887. doi: 10.3390/jcm13164887. J Clin Med. 2024. PMID: 39201028 Free PMC article. Review.
-
Association Between Sarcopenia and Adverse Events Following Transcatheter Aortic Valve Implantation.CJC Open. 2021 Sep 16;4(2):173-179. doi: 10.1016/j.cjco.2021.09.012. eCollection 2022 Feb. CJC Open. 2021. PMID: 35198934 Free PMC article.
-
Prognosis after surgical replacement with a bioprosthetic aortic valve in patients with severe symptomatic aortic stenosis: systematic review of observational studies.BMJ. 2016 Sep 28;354:i5065. doi: 10.1136/bmj.i5065. BMJ. 2016. PMID: 27683072 Free PMC article.
-
Mid-term follow-up after suture-less aortic heart valve implantation.J Thorac Dis. 2018 Nov;10(11):6128-6136. doi: 10.21037/jtd.2018.10.10. J Thorac Dis. 2018. PMID: 30622784 Free PMC article.
-
Rapid deployment technology versus conventional sutured bioprostheses in aortic valve replacement.J Card Surg. 2022 Mar;37(3):640-655. doi: 10.1111/jocs.16223. Epub 2022 Jan 14. J Card Surg. 2022. PMID: 35028981 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous