Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2015 Mar;17(3):379-87.
doi: 10.1093/europace/euu273. Epub 2014 Nov 5.

Safety and efficacy outcomes in patients undergoing pulmonary vein isolation with second-generation cryoballoon†

Affiliations

Safety and efficacy outcomes in patients undergoing pulmonary vein isolation with second-generation cryoballoon†

Kudret Aytemir et al. Europace. 2015 Mar.

Abstract

Aims: The second-generation cryoballoon (Arctic Front Advance™) (Arc-Adv-CB) has a redesigned injection system which distributes the refrigerant homogenously to the frontal balloon surface. The aim of this study was to compare the efficacy and safety of the Arc-Adv-CB and its predecessor (Arctic Front™) (Arc-CB) in patients who underwent pulmonary vein isolation (PVI) for atrial fibrillation (AF).

Methods and results: Three hundred and six patients (55.35 ± 10.60 years, 47.05% male) were included in the study. A total of 1205 pulmonary veins were attempted for PVI with either Arc-CB or Arc-Adv-CB. The follow-up durations were 30 (23-38) and 10 (8-13) months in Arc-CB and Arc-Adv-CB groups, respectively (P < 0.001). When the blanking period was considered, freedom from AF after a single ablation procedure was 68.53 and 90.83% in patients undergoing PVI with Arc-CB and Arc-Adv-CB, respectively. The most frequent complication was transient phrenic nerve palsy (PNP) which occurred in five(2.54%) and nine(8.26%) of patients undergoing PVI with Arc-CB and Arc-Adv-CB, respectively (P = 0.040). Left atrial (LA) diameter (hazard ratio, HR: 3.552, 95% CI: 2.034-6.201, P < 0.001), smoking history (HR:1.643, 95% CI: 1.011-2.671, P = 0.045), persistent AF (HR:1.725, 95% CI: 1.021-2.915, P = 0.041), duration of AF (HR:1.039, 95% CI: 1.000-1.080, P = 0.047), and early AF recurrence (HR:2.399, 95% CI: 1.443-3.989, P < 0.001) were associated with increased late AF recurrence. On the other hand, intraprocedural vagal reactions (HR: 0.550, 95% CI: 0.331-0.915, P = 0.021) and Arc-Adv-CB use (HR: 0.441, 95% CI: 0.225-0.866, P = 0.017) were associated with lower late AF recurrence. Left atrial diameter (HR: 3.072, 95% CI: 1.646-5.732, P < 0.001), early AF recurrence (HR: 1.906, 95% CI: 1.103-3.291, P = 0.021), and Arc-Adv-CB use (HR: 0.472, 95% CI: 0.239-0.931, P = 0.030) were independent predictors for late AF recurrence.

Conclusion: Our study has shown that Arc-Adv-CB use is associated with lower late AF recurrences at the cost of an increased risk for PNP.

Keywords: Ablation; Atrial fibrillation; Cryoballoon; Efficacy; Safety.