Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpflug imaging devices
- PMID: 25410175
- DOI: 10.5935/0004-2749.20140060
Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpflug imaging devices
Abstract
Purpose: To compare the ocular anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter, as measured with two different Scheimpflug imaging devices.
Methods: This transversal study included 80 right eyes from 80 subjects aged from 20 to 40 years. Their spherical equivalents ranged from -4.25 to +1.00 diopters (D). Each eye's anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter, were measured for far vision using both the Galilei G4 (double Scheimpflug camera) and the Pentacam HR (single Scheimpflug camera) systems.
Results: Mean anterior chamber depths were calculated as 3.12 ± 0.23 mm and 3.19 ± 0.24 mm when measured with the Galilei G4 and the Pentacam HR, respectively. The mean white-to-white distance measured was 11.84 ± 0.31 mm and 11.90 ± 0.43 mm when measured with the Galilei G4 and the Pentacam HR, respectively. Mean pupil diameters were measured as 3.22 ± 0.58 mm and 3.22 ± 0.52 mm when measured with the Galilei G4 and the Pentacam HR, respectively. Finally, the mean anterior chamber angle was 34.30 ± 2.86 degrees when it was measured with the Galilei G4, and 39.26 ± 2.85 degrees when measured with the Pentacam HR. A comparative analysis revealed that the Galilei G4 yielded a significantly lower (P<0.05) measurement for the anterior chamber depth, anterior chamber angle, and pupil diameter, than the Pentacam HR system. Comparable values (P>0.05) for both devices were obtained for the white-to-white distance measurements.
Conclusion: The Galilei G4 and Pentacam HR Scheimpflug systems cannot be used interchangeably because they produce significant measurement differences.
Similar articles
-
Comparison between Two Scheimpflug Anterior Segment Analyzers.J Ophthalmic Vis Res. 2017 Jan-Mar;12(1):23-29. doi: 10.4103/jovr.jovr_104_16. J Ophthalmic Vis Res. 2017. PMID: 28299003 Free PMC article.
-
Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II.Optometry. 2010 Jan;81(1):35-9. doi: 10.1016/j.optm.2009.04.100. Optometry. 2010. PMID: 20004876
-
Changes in the anterior chamber during accommodation assessed with a Scheimpflug system.J Cataract Refract Surg. 2014 Nov;40(11):1790-7. doi: 10.1016/j.jcrs.2014.02.043. Epub 2014 Sep 10. J Cataract Refract Surg. 2014. PMID: 25217074
-
[Pentacam--the new way for anterior eye segment imaging and mapping].Klin Oczna. 2005;107(10-12):728-31. Klin Oczna. 2005. PMID: 16619832 Review. Polish.
-
Galilei Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part II.Med Hypothesis Discov Innov Ophthalmol. 2019 Fall;8(3):204-218. Med Hypothesis Discov Innov Ophthalmol. 2019. PMID: 31598521 Free PMC article. Review.
Cited by
-
Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.Int J Ophthalmol. 2016 Jul 18;9(7):1057-65. doi: 10.18240/ijo.2016.07.20. eCollection 2016. Int J Ophthalmol. 2016. PMID: 27500117 Free PMC article. Review.
-
Repeatability of pupil diameter measurements using three different topography devices.PLoS One. 2023 Aug 18;18(8):e0290417. doi: 10.1371/journal.pone.0290417. eCollection 2023. PLoS One. 2023. PMID: 37594948 Free PMC article.
-
A comparison of Scansys and Sirius tomography in healthy eyes.BMC Ophthalmol. 2024 Mar 27;24(1):138. doi: 10.1186/s12886-024-03389-7. BMC Ophthalmol. 2024. PMID: 38539111 Free PMC article.
-
Comparison of Anterior Corneal Aberrometry, Keratometry and Pupil Size with Scheimpflug Tomography and Ray Tracing Aberrometer.Vision (Basel). 2022 Mar 18;6(1):18. doi: 10.3390/vision6010018. Vision (Basel). 2022. PMID: 35324603 Free PMC article.
-
Comparison between Two Scheimpflug Anterior Segment Analyzers.J Ophthalmic Vis Res. 2017 Jan-Mar;12(1):23-29. doi: 10.4103/jovr.jovr_104_16. J Ophthalmic Vis Res. 2017. PMID: 28299003 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources