Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer
- PMID: 25431133
- DOI: 10.3109/00365513.2014.981854
Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer
Abstract
Background: Blood gas analyzers (BGAs) are important in assessing and monitoring critically ill patients. However, the random use of BGAs to measure blood gases, electrolytes and metabolites increases the variability in test results. Therefore, this study aimed to investigate the correlation of blood gas, electrolyte and metabolite results measured with two BGAs and a core laboratory analyzer.
Methods: A total of 40 arterial blood gas samples were analyzed with two BGAs [(Nova Stat Profile Critical Care Xpress (Nova Biomedical, Waltham, MA, USA) and Siemens Rapidlab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA)) and a core laboratory analyzer [Olympus AU 2700 autoanalyzer (Beckman-Coulter, Inc., Fullerton, CA, USA)]. The results of pH, pCO₂, pO₂, SO₂, sodium (Na⁺), potassium (K⁺), calcium (Ca⁺²), chloride (Cl⁻), glucose, and lactate were compared by Passing-Bablok regression analysis and Bland-Altman plots.
Results: The present study showed that there was negligible variability of blood gases (pCO₂, pO₂, SO₂), K⁺ and lactate values between the blood gas and core laboratory analyzers. However, the differences in pH were modest, while Na⁺, Cl⁻, Ca²⁺ and glucose showed poor correlation according to the concordance correlation coefficient.
Conclusions: BGAs and core laboratory autoanalyzer demonstrated variable performances and not all tests met minimum performance goals. It is important that clinicians and laboratories are aware of the limitations of their assays.
Keywords: Blood gas analyzer; electrolytes; interchangeability; metabolites; method comparison.
Similar articles
-
Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers.Clin Chem Lab Med. 2011 Jul;49(7):1187-91. doi: 10.1515/CCLM.2011.185. Epub 2011 Apr 20. Clin Chem Lab Med. 2011. PMID: 21504373
-
Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting.Clin Chem Lab Med. 2016 Apr;54(4):585-94. doi: 10.1515/cclm-2015-0592. Clin Chem Lab Med. 2016. PMID: 26457784
-
Comparison of Enterprise Point-of-Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood.Vet Clin Pathol. 2018 Sep;47(3):415-424. doi: 10.1111/vcp.12635. Epub 2018 Jul 9. Vet Clin Pathol. 2018. PMID: 29989207
-
Compatibility levels between blood gas analysis and central laboratory hemoglobin and electrolyte tests in pediatric patients: A single-center experience.Paediatr Anaesth. 2023 Feb;33(2):107-113. doi: 10.1111/pan.14567. Epub 2022 Oct 10. Paediatr Anaesth. 2023. PMID: 36178754 Review.
-
Point-of-care blood gas and electrolyte analysis in rabbits.Vet Clin North Am Exot Anim Pract. 2013 Jan;16(1):175-95. doi: 10.1016/j.cvex.2012.10.005. Epub 2012 Dec 12. Vet Clin North Am Exot Anim Pract. 2013. PMID: 23347543 Review.
Cited by
-
A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease.Biochem Med (Zagreb). 2020 Oct 15;30(3):030702. doi: 10.11613/BM.2020.030702. Epub 2020 Aug 5. Biochem Med (Zagreb). 2020. PMID: 32774124 Free PMC article.
-
Comparison of electrolyte and glucose levels measured by a blood gas analyzer and an automated biochemistry analyzer among hospitalized patients.J Clin Lab Anal. 2020 Jul;34(7):e23291. doi: 10.1002/jcla.23291. Epub 2020 Mar 9. J Clin Lab Anal. 2020. PMID: 32147884 Free PMC article.
-
Pseudohyperkalemia Associated With Leukemia.Cureus. 2022 Apr 9;14(4):e23978. doi: 10.7759/cureus.23978. eCollection 2022 Apr. Cureus. 2022. PMID: 35541290 Free PMC article.
-
Plasma sodium measurements by direct ion selective methods in laboratory and point of care may not be clinically interchangeable.J Clin Monit Comput. 2017 Oct;31(5):1103-1109. doi: 10.1007/s10877-016-9938-1. Epub 2016 Sep 28. J Clin Monit Comput. 2017. PMID: 27683272
-
Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions.PLoS One. 2018 Nov 12;13(11):e0206707. doi: 10.1371/journal.pone.0206707. eCollection 2018. PLoS One. 2018. PMID: 30418977 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous