Institutional patient-specific IMRT QA does not predict unacceptable plan delivery
Abstract
Purpose: To determine whether in-house patient-specific intensity modulated radiation therapy quality assurance (IMRT QA) results predict Imaging and Radiation Oncology Core (IROC)-Houston phantom results.
Methods and materials: IROC Houston's IMRT head and neck phantoms have been irradiated by numerous institutions as part of clinical trial credentialing. We retrospectively compared these phantom results with those of in-house IMRT QA (following the institution's clinical process) for 855 irradiations performed between 2003 and 2013. The sensitivity and specificity of IMRT QA to detect unacceptable or acceptable plans were determined relative to the IROC Houston phantom results. Additional analyses evaluated specific IMRT QA dosimeters and analysis methods.
Results: IMRT QA universally showed poor sensitivity relative to the head and neck phantom, that is, poor ability to predict a failing IROC Houston phantom result. Depending on how the IMRT QA results were interpreted, overall sensitivity ranged from 2% to 18%. For different IMRT QA methods, sensitivity ranged from 3% to 54%. Although the observed sensitivity was particularly poor at clinical thresholds (eg 3% dose difference or 90% of pixels passing gamma), receiver operator characteristic analysis indicated that no threshold showed good sensitivity and specificity for the devices evaluated.
Conclusions: IMRT QA is not a reasonable replacement for a credentialing phantom. Moreover, the particularly poor agreement between IMRT QA and the IROC Houston phantoms highlights surprising inconsistency in the QA process.
Copyright © 2014 Elsevier Inc. All rights reserved.
Conflict of interest statement
Figures
References
-
- Das IJ, Cheng CW, Chopra KL, Mitra RK, Srivastava SP, Glatstein E. Intensity-modulated radiation therapy dose prescription, recording, and delivery: patterns of variability among institutions and treatment planning systems. J Natl Cancer Inst. 2008;100:300–307. - PubMed
-
- Peters LF, O’Sullivan B, Giralt J, et al. Critical impact of radiotherapy protocol compliance and quality in the treatment of advanced head and neck cancer: results from TROG 02. 02. J Clin Oncol. 2010;28:2996–3001. - PubMed
-
- Bekelman J, Deye J, Vikram B, et al. Redesigning radiotherapy quality assurance: opportunities to develop an efficient, evidence-based system to support clinical trials – report of the National Cancer Institute work group on radiotherapy quality assurance. Int J Radiat Oncol Biol Phys. 2012;83(3):782–790. - PMC - PubMed
-
- Molineu A, Followill DS, Balter P, et al. Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group. Int J Radiat Oncol Biol Phys. 2005;63:577–583. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical