. 2015 Feb;12(2):330-7.
doi: 10.1016/j.hrthm.2014.10.028. Epub 2014 Oct 29.

Effect of daily remote monitoring on pacemaker longevity: a retrospective analysis

Affiliations

Effect of daily remote monitoring on pacemaker longevity: a retrospective analysis

Renato Pietro Ricci et al. Heart Rhythm. 2015 Feb.

Abstract

Background: Energy demand of remote monitoring in cardiac implantable electronic devices has never been investigated. Biotronik Home Monitoring (HM) is characterized by daily transmissions that may affect longevity.

Objective: The aim of the study was to retrospectively compare longevity of a specific dual-chamber pacemaker model in patients with HM on and patients with HM off.

Methods: Hospital files of 201 patients (mean age 87 ± 10 years, 78 men) who had received a Biotronik Cylos DR-T pacemaker between April 2006 and May 2010 for standard indication were reviewed. In 134 patients (67%), HM was activated. The primary end point was device replacement due to battery depletion.

Results: The median follow-up period was 56.4 months (interquartile range 41.8-65.2 months). The estimated device longevity was 71.1 months (95% confidence interval [CI] 69.1-72.3 months) in the HM-on group and 60.4 months (CI 55.9-65.1 months) in the HM-off group (P < .0001). The frequency of inhospital visits with significant device reprogramming was higher in the HM-on group than in the HM-off group (33.3% vs 25.0%, respectively; P = .03). Lower ventricular pulse amplitude (2.3 ± 0.4 V vs 2.7 ± 0.5 V; P < .0001) and pacing percentage (49% ± 38% vs 64% ± 38%; P = .02), both calculated as time-weighted averages, were observed with HM on as compared with HM off. Patient attrition was significantly lower in the HM-on group (9.7%; 95% CI 3.0%-28.7%) than in the HM-off group (45.6%; 95% CI 30.3%-64.3%) (P < .0001).

Conclusion: In normal practice, energy demand of HM, if present, was overshadowed by programming optimization likely favored by continuous monitoring. Pacemakers controlled remotely with HM showed an 11-month longer longevity. Patient retention was superior.

Keywords: Battery; Device programming; Home Monitoring; Longevity; Pacemaker; Remote monitoring; Telemedicine.