Image guided particle therapy in CNAO room 2: implementation and clinical validation
- PMID: 25455440
- DOI: 10.1016/j.ejmp.2014.10.075
Image guided particle therapy in CNAO room 2: implementation and clinical validation
Abstract
In this contribution we describe the implementation of a novel solution for image guided particle therapy, designed to ensure the maximal accuracy in patient setup. The presented system is installed in the central treatment room at Centro Nazionale di Adroterapia Oncologica (CNAO, Italy), featuring two fixed beam lines (horizontal and vertical) for proton and carbon ion therapy. Treatment geometry verification is based on robotic in-room imaging acquisitions, allowing for 2D/3D registration from double planar kV-images or 3D/3D alignment from cone beam image reconstruction. The calculated six degrees-of-freedom correction vector is transferred to the robotic patient positioning system, thus yielding automated setup error compensation. Sub-millimetre scale residual errors were measured in absolute positioning of rigid phantoms, in agreement with optical- and laser-based assessment. Sub-millimetre and sub-degree positioning accuracy was achieved when simulating setup errors with anthropomorphic head, thorax and pelvis phantoms. The in-house design and development allowed a high level of system customization, capable of replicating the clinical performance of commercially available products, as reported with preliminary clinical results in 10 patients.
Keywords: Accuracy measurement; CBCT; IGRT; Image registration; Optical imaging; Robotic imaging.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
A new system for in-room image guidance in particle therapy at CNAO.Phys Med. 2023 Oct;114:103162. doi: 10.1016/j.ejmp.2023.103162. Epub 2023 Oct 9. Phys Med. 2023. PMID: 37820507
-
An accuracy assessment of different rigid body image registration methods and robotic couch positional corrections using a novel phantom.Med Phys. 2013 Mar;40(3):031701. doi: 10.1118/1.4789490. Med Phys. 2013. PMID: 23464296
-
A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance.Br J Radiol. 2017 Nov;90(1079):20170266. doi: 10.1259/bjr.20170266. Epub 2017 Oct 9. Br J Radiol. 2017. PMID: 28830239 Free PMC article.
-
Dynamic targeting image-guided radiotherapy.Med Dosim. 2006 Summer;31(2):113-25. doi: 10.1016/j.meddos.2005.12.014. Med Dosim. 2006. PMID: 16690452 Review.
-
Current state and future applications of radiological image guidance for particle therapy.Med Phys. 2018 Nov;45(11):e1086-e1095. doi: 10.1002/mp.12744. Med Phys. 2018. PMID: 30421805 Review.
Cited by
-
Image-based shading correction for narrow-FOV truncated pelvic CBCT with deep convolutional neural networks and transfer learning.Med Phys. 2021 Nov;48(11):7112-7126. doi: 10.1002/mp.15282. Epub 2021 Oct 26. Med Phys. 2021. PMID: 34636429 Free PMC article.
-
Feasibility Study on Cardiac Arrhythmia Ablation Using High-Energy Heavy Ion Beams.Sci Rep. 2016 Dec 20;6:38895. doi: 10.1038/srep38895. Sci Rep. 2016. PMID: 27996023 Free PMC article.
-
Dosimetric Impact of Inter-Fraction Anatomical Changes in Carbon Ion Boost Treatment for High-Risk Prostate Cancer (AIRC IG 14300).Front Oncol. 2021 Sep 28;11:740661. doi: 10.3389/fonc.2021.740661. eCollection 2021. Front Oncol. 2021. PMID: 34650922 Free PMC article.
-
Topogram and 3DCT geometry calibration for image-guided proton therapy with in-room CT-on-rails.Phys Imaging Radiat Oncol. 2025 Jun 24;35:100799. doi: 10.1016/j.phro.2025.100799. eCollection 2025 Jul. Phys Imaging Radiat Oncol. 2025. PMID: 40678015 Free PMC article.
-
Time-resolved MRI for off-line treatment robustness evaluation in carbon-ion radiotherapy of pancreatic cancer.Med Phys. 2022 Apr;49(4):2386-2395. doi: 10.1002/mp.15510. Epub 2022 Feb 17. Med Phys. 2022. PMID: 35124811 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources