Unexpected triggering of the dialysate blood leak detector by haemolysis
- PMID: 25468361
- DOI: 10.1179/2295333714Y.0000000105
Unexpected triggering of the dialysate blood leak detector by haemolysis
Abstract
This case showed that it is possible for haemoglobin released by haemolysis in the extracorporeal circuit to pass through a high permeability (albumin retaining) dialyser membrane. In the incident described, the blood leak detector of the dialysis machine was activated before the patient became symptomatic. Haemolysis should be considered as a possible cause of blood leak alarms during dialysis with high flux membranes.
Keywords: Blood leak alarm; Haemodialysis,; Haemolysis,; High-permeability membrane,.