Comparison of image quality of shoulder CT arthrography conducted using 120 kVp and 140 kVp protocols
- PMID: 25469085
- PMCID: PMC4248629
- DOI: 10.3348/kjr.2014.15.6.739
Comparison of image quality of shoulder CT arthrography conducted using 120 kVp and 140 kVp protocols
Abstract
Objective: To compare the image quality of shoulder CT arthrography performed using 120 kVp and 140 kVp protocols.
Materials and methods: Fifty-four CT examinations were prospectively included. CT scans were performed on each patient at 120 kVp and 140 kVp; other scanning parameters were kept constant. Image qualities were qualitatively and quantitatively compared with respect to noise, contrast, and diagnostic acceptability. Diagnostic acceptabilities were graded using a one to five scale as follows: 1, suboptimal; 2, below average; 3, acceptable; 4, above average; and 5, superior. Radiation doses were also compared.
Results: Contrast was better at 120 kVp, but noise was greater. No significant differences were observed between the 120 kVp and 140 kVp protocols in terms of diagnostic acceptability, signal-to-noise ratio, or contrast-to-noise ratio. Lowering tube voltage from 140 kVp to 120 kVp reduced the radiation dose by 33%.
Conclusion: The use of 120 kVp during shoulder CT arthrography reduces radiation dose versus 140 kVp without significant loss of image quality.
Keywords: Image quality; Radiation dose; Shoulder CT arthrography; Tube voltage.
Figures


Similar articles
-
Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study.Skeletal Radiol. 2019 Aug;48(8):1261-1268. doi: 10.1007/s00256-019-3150-6. Epub 2019 Feb 6. Skeletal Radiol. 2019. PMID: 30725159
-
Effect of reduced x-ray tube voltage, low iodine concentration contrast medium, and sinogram-affirmed iterative reconstruction on image quality and radiation dose at coronary CT angiography: results of the prospective multicenter REALISE trial.J Cardiovasc Comput Tomogr. 2015 May-Jun;9(3):215-24. doi: 10.1016/j.jcct.2015.01.010. Epub 2015 Jan 22. J Cardiovasc Comput Tomogr. 2015. PMID: 25843243 Clinical Trial.
-
Image quality and radiation dose of pulmonary CT angiography performed using 100 and 120 kVp.AJR Am J Roentgenol. 2012 Nov;199(5):990-6. doi: 10.2214/AJR.11.8208. AJR Am J Roentgenol. 2012. PMID: 23096170
-
Low tube voltage computed tomography urography using low-concentration contrast media: Comparison of image quality in conventional computed tomography urography.Eur J Radiol. 2015 Dec;84(12):2454-63. doi: 10.1016/j.ejrad.2015.09.010. Epub 2015 Sep 10. Eur J Radiol. 2015. PMID: 26388465 Clinical Trial.
-
Low-dose abdominal CT protocols with a tube voltage setting of 100 kVp or 80 kVp: Performance of radiation dose reduction and influence on visual contrast.Clin Radiol. 2014 Aug;69(8):804-11. doi: 10.1016/j.crad.2014.03.013. Epub 2014 Jun 3. Clin Radiol. 2014. PMID: 24907185 Clinical Trial.
Cited by
-
C-arm flat-panel CT arthrography of the shoulder: Radiation dose considerations and preliminary data on diagnostic performance.Eur Radiol. 2017 Feb;27(2):454-463. doi: 10.1007/s00330-016-4382-7. Epub 2016 May 24. Eur Radiol. 2017. PMID: 27221562
-
Influence of joint volume on range of motion after arthroscopic rotator cuff repair.BMC Musculoskelet Disord. 2023 Mar 17;24(1):201. doi: 10.1186/s12891-023-06306-z. BMC Musculoskelet Disord. 2023. PMID: 36932406 Free PMC article.
-
Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study.Skeletal Radiol. 2019 Aug;48(8):1261-1268. doi: 10.1007/s00256-019-3150-6. Epub 2019 Feb 6. Skeletal Radiol. 2019. PMID: 30725159
References
-
- Lecouvet FE, Simoni P, Koutaïssoff S, Vande Berg BC, Malghem J, Dubuc JE. Multidetector spiral CT arthrography of the shoulder. Clinical applications and limits, with MR arthrography and arthroscopic correlations. Eur J Radiol. 2008;68:120–136. - PubMed
-
- Oh JH, Kim JY, Choi JA, Kim WS. Effectiveness of multidetector computed tomography arthrography for the diagnosis of shoulder pathology: comparison with magnetic resonance imaging with arthroscopic correlation. J Shoulder Elbow Surg. 2010;19:14–20. - PubMed
-
- Kim YJ, Choi JA, Oh JH, Hwang SI, Hong SH, Kang HS. Superior labral anteroposterior tears: accuracy and interobserver reliability of multidetector CT arthrography for diagnosis. Radiology. 2011;260:207–215. - PubMed
-
- De Filippo M, Araoz PA, Pogliacomi F, Castagna A, Petriccioli D, Sverzellati N, et al. Recurrent superior labral anterior-to-posterior tears after surgery: detection and grading with CT arthrography. Radiology. 2009;252:781–788. - PubMed
-
- Choi JY, Kim SH, Yoo HJ, Shin SH, Oh JH, Baek GH, et al. Superior labral anterior-to-posterior lesions: comparison of external rotation and active supination CT arthrography with neutral CT arthrography. Radiology. 2012;263:199–205. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical