Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2015 Mar;38(3):302-5.
doi: 10.1111/pace.12552. Epub 2014 Dec 2.

Removal of active-fixation coronary sinus leads using a mechanical rotation extraction device

Affiliations

Removal of active-fixation coronary sinus leads using a mechanical rotation extraction device

Alexander Kypta et al. Pacing Clin Electrophysiol. 2015 Mar.

Abstract

Background: Active fixation coronary sinus (CS) leads are widely used in cardiac resynchronization therapy (CRT). Due to their low dislodgement rates they are an attractive option for implanters. However, extraction of active fixation leads is a complex procedure that bears potential fatal risks for patients.

Methods: We analyzed all patients undergoing StarFix® (Medtronic Inc., Minneapolis, MN, USA) extraction because of severe infection at our institution. Indication for extraction was severe device infection. Procedural outcomes as well as patient characteristics are reported. For removal of the leads a 9-Fr Evolution® (Cook Intravascular Inc., Leechburg, PA, USA) mechanical rotation extraction sheath was used.

Results: Between 2011 and 2014, six patients underwent extraction of StarFix® leads. One patient was female, whereas the others were males (mean age 64 ± 12 years). Implant duration was 46.5 ± 8.2 months before removal. All leads could be successfully extracted totally with the Evolution®. In one patient (16.6%) pericardial tamponade occurred immediately requiring surgical intervention. After establishment of cardiopulmonary bypass a 5-mm sharp cut in the lateral vein of the CS could be identified. It was fixed and the patient survived. On all leads, significant tissue growth between the fixation lobes could be observed.

Conclusion: As illustrated by our cases, Attain StarFix® leads implanted years ago can be extracted with the help of a mechanical extraction sheath. However, this procedure bears potential risks and should only be performed with a cardiac surgery standby. If clinicians evaluate the implantation of Attain StarFix® leads in patients who are young or at high risk for device infection, they should be aware of these findings.

Keywords: active fixation leads; device infection; lead extraction; resynchronization therapy.