Comparison of different surface modifications for titanium implants installed into the goat iliac crest
- PMID: 25486997
- DOI: 10.1111/clr.12529
Comparison of different surface modifications for titanium implants installed into the goat iliac crest
Abstract
Objectives: This in vivo study with implants installed in the goat iliac crest was performed to determine whether the biological and mechanical properties of the bone-to-implant interface are influenced by (i) the type of implant anchorage (i.e., mono- vs. bicortical placement), and (ii) the presence of a bioactive hydroxyapatite (HA) or composite HA/bioactive glass (BG) coatings.
Materials and methods: A total of 96 titanium (Ti) implants w/- coatings (Ti, Ti-HA & Ti-HABG; n = 8) were mono- or bicortically placed in the iliac crest of eight goats. At installation and after 4 weeks, implant stability was determined using insertion and removal torque testing (ITQ & RTQ). The peri-implant bone response was histologically and histomorphometrically evaluated by means of bone-to-implant contact (BIC%) and bone area (BA%).
Results: Monocortical implants demonstrated significantly lower RTQ values in comparison to ITQ values, whereas for bicortical implant placement RTQ and ITQ were similar. Further, mean RTQ values for monocortical implants were significantly lower in comparison to bicortical implants. Histomorphometrical evaluation demonstrated higher BIC% and BA% for bicortical implants compared to monocortical implants. For bicortical implants, BA% in the inner peri-implant region (0-500 μm) was significantly higher compared to the middle (500-1000 μm) and outer (1000-1500 μm) region. Also, a significant correlation was observed for monocortical implants between RTQ and BIC% and BA%. For surface modifications, no significant differences were found in ITQ and RTQ, for neither mono- nor bicortical implants. Histomorphometrically, HABG-coated implants demonstrated significantly higher BIC% compared to GAE surfaces for both mono- and bicortical implants. Bicortical HA-coated implants revealed significant higher BA% in the inner peri-implant region (0-500 μm) in comparison to bicortical GAE implants.
Conclusions: This study demonstrated that bicortical implant placement beneficially affects implant stability during the early phase of osseointegration. A significant correlation between removal torque and bone-to-implant contact and bone area for monocortical implants was observed, but not for bicortical implants. Therefore, histomorphometrical data should be interpreted with caution to predict the biomechanical implant fixation of bone implants over time. Regarding surface modifications, in the present implantation model, the addition of BG to an RF magnetron sputtered HA coating enhanced the biological behavior of the coating compared to grit-blasted/acid-etched implants.
© 2014 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.
Similar articles
-
In vivo evaluation of bioactive glass-based coatings on dental implants in a dog implantation model.Clin Oral Implants Res. 2014 Jan;25(1):21-8. doi: 10.1111/clr.12060. Epub 2012 Oct 19. Clin Oral Implants Res. 2014. PMID: 23078340
-
Evaluation of primary and secondary stability of titanium implants using different surgical techniques.Clin Oral Implants Res. 2014 Apr;25(4):487-92. doi: 10.1111/clr.12180. Epub 2013 May 3. Clin Oral Implants Res. 2014. PMID: 23638908
-
In vivo bone response and mechanical evaluation of electrosprayed CaP nanoparticle coatings using the iliac crest of goats as an implantation model.Acta Biomater. 2010 Jun;6(6):2227-36. doi: 10.1016/j.actbio.2009.11.030. Epub 2009 Nov 26. Acta Biomater. 2010. PMID: 19944782
-
Comparison of the Use of Titanium-Zirconium Alloy and Titanium Alloy in Dental Implants: A Systematic Review and Meta-Analysis.J Oral Implantol. 2018 Aug;44(4):305-312. doi: 10.1563/aaid-joi-D-17-00233. Epub 2018 Feb 19. J Oral Implantol. 2018. PMID: 29457740
-
Strontium in the bone-implant interface.Dan Med Bull. 2011 May;58(5):B4286. Dan Med Bull. 2011. PMID: 21535993 Review.
Cited by
-
The Influence of Nanostructured Hydroxyapatite Surface in the Early Stages of Osseointegration: A Multiparameter Animal Study in Low-Density Bone.Int J Nanomedicine. 2020 Nov 10;15:8803-8817. doi: 10.2147/IJN.S280957. eCollection 2020. Int J Nanomedicine. 2020. PMID: 33204089 Free PMC article.
-
Evaluation of Clinical Performance of TiNi-Based Implants Used in Chest Wall Repair after Resection for Malignant Tumors.J Funct Biomater. 2021 Nov 11;12(4):60. doi: 10.3390/jfb12040060. J Funct Biomater. 2021. PMID: 34842727 Free PMC article.
-
Bioactive Surfaces vs. Conventional Surfaces in Titanium Dental Implants: A Comparative Systematic Review.J Clin Med. 2020 Jun 29;9(7):2047. doi: 10.3390/jcm9072047. J Clin Med. 2020. PMID: 32610687 Free PMC article. Review.
-
Bioactive Glass and Silicate-Based Ceramic Coatings on Metallic Implants: Open Challenge or Outdated Topic?Materials (Basel). 2019 Sep 10;12(18):2929. doi: 10.3390/ma12182929. Materials (Basel). 2019. PMID: 31510062 Free PMC article. Review.
-
Bioactive Glasses and Glass-Ceramics for Healthcare Applications in Bone Regeneration and Tissue Engineering.Materials (Basel). 2018 Dec 12;11(12):2530. doi: 10.3390/ma11122530. Materials (Basel). 2018. PMID: 30545136 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous