Observational Study
. 2014 Nov 15;177(1):182-6.
doi: 10.1016/j.ijcard.2014.09.026. Epub 2014 Sep 28.

Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction

Affiliations
Observational Study

Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction

Sally Aldous et al. Int J Cardiol. .

Abstract

Objectives: The aim of this study is to compare a new improved point of care cardiac troponin assay (new POC-cTnI) with 1. its predecessor (old POC-cTnI) and 2. a high sensitivity assay (hs-cTnI) for the diagnosis of acute myocardial infarction (AMI) and for major adverse cardiac events (MACE) by 30 days.

Methods: This is a single centre observational study, set in Christchurch Hospital, New Zealand. Patients presenting to the emergency department with non-traumatic chest pain underwent blood sampling at 0 h and 2h post presentation for analysis with the 3 cTnI assays for the outcome of AMI and for analysis using an accelerated diagnostic protocol (ADP-normal 2h troponins, normal electrocardiograms and Thrombolysis In Myocardial Infarction (TIMI) score of 0 or ≤ 1) for 30 day MACE.

Results: Of 962 patients, 220 (22.9%) had AMI. Old POC-cTnI was least sensitive at 70.0% (65.4-73.9%) by 2h (p<0.001). New POC-cTnI, sensitivity 93.6% (89.9-96.2%) had similar sensitivity to hs-cTnI, sensitivity 95.0% (91.5-97.3%) (p = 0.508). There were 231 (24.0%) patients with 30 day MACE. When used as part of the ADP, all assays had 100% (98.0-100%) sensitivity using TIMI = 0. Sensitivities of new POC-cTnI ADP, 98.3% (95.4-99.4%), old POC-cTnI, 96.5% (93.2-98.4%) and hs-cTnI, 98.7% (96.0-99.7%) were similar (p = 0.063-0.375) using TIMI ≤ 1.

Conclusions: A new POC-cTnI has improved sensitivity for AMI and MACE compared with its predecessor and comparable sensitivity to a high sensitivity assay. Now that sensitivities of the POC assay are improved, the new assay may be a useful alternative to central laboratory assays when rapid turn-around times are not possible.

Keywords: Accelerated diagnostic protocol; Acute myocardial infarction; High sensitivity troponin; Major adverse cardiac events; Point of care troponin.

PubMed Disclaimer

Publication types

LinkOut - more resources