Abstract
Background: Postoperative periprosthetic femur fractures are an increasing concern after primary total hip arthroplasty (THA). Identifying and understanding predisposing factors are important to mitigating future risk. Femoral stem design may be one such factor.
Questions/purposes: The goals of our study were to compare the (1) frequency of periprosthetic femur fracture and implant survivorship; (2) time to fracture in those patients who experienced periprosthetic femur fracture; and (3) predictive risk factors for periprosthetic femur fracture between a unique stem design with an exaggerated proximal taper angle and other contemporary cementless, proximally fixed, tapered stems.
Methods: We reviewed all hips in which a femoral hip component with a uniquely exaggerated proximal taper angle (ProxiLock) was implanted during primary THA at a single academic institution. That group of patients was compared with a cohort of patients who underwent primary THA during the same time interval (1995-2008) in which any other cementless, proximally fixed, tapered stem design was used. The two groups differed somewhat in terms of sex, age, and body mass index, although these differences were of unclear clinical significance. During the study, 3964 primary THAs were performed using six different designs of cementless, proximally fixed, tapered femoral hip prostheses. There were 736 stems in the ProxiLock (PL) patient group and 3228 stems in the non-ProxiLock (non-PL) group. In general, the stem highlighted in this study became the routine cementless stem used for primary THA for three arthroplasty surgeons without specific patient or radiographic indications. Periprosthetic fractures were identified within each group. The incidence, timing, type, and treatment required for each fracture were analyzed. The Kaplan-Meier method was used to determine study patient survival free of any postoperative fracture. Radiographs and the electronic medical record of each patient who sustained a fracture were reviewed. Followup was comparable between groups at all time points.
Results: The Kaplan-Meier estimate for fracture-free patient survival was worse in the PL group at all time points with survival of 98.4% (range, 97.4%-99.3%), 97.1% (range, 95.9%-98.3%), 95.4% (range, 93.8%-97.0%), and 92.6% (range, 89.6%-95.3%) at 30 days, 1 year, 5 years, and 10 years, respectively, for the PL patient group compared with 99.8% (range, 99.7%-99.9%), 99.6% (range, 99.3%-99.8%), 99.3% (range, 99.0%-99.6%), and 98.4% (range, 97.5%-99.1%) in the non-PL patient group (p < 0.001). Patients in the PL group had increased cumulative probability of both early and late fractures with cumulative probabilities of fracture of 2.5% (range, 1.3%-3.6%) at 90 days and 7.4% (range, 4.7%-10.4%) at 10 years compared with probabilities of 0.3% (range, 0.1%-0.5%) at 90 days and 1.6% (range, 0.8%-2.5%) at 10 years in the non-PL group (p < 0.001). Patients in the PL group had an increased risk of postoperative periprosthetic femur fracture (hazard ratio [HR], 5.6; 95% confidence interval [CI], 3.4-9.1; p < 0.001); fracture requiring reoperation (HR, 8.4; 95% CI, 4.4-15.9); p < 0.001); and fracture requiring stem revision (HR, 9.1; 95% CI, 4.5-18.5; p < 0.001). Age older than 60 years was also a risk factor for fracture (HR, 3.7; 95% CI, 2.1-6.4), but sex, body mass index, and preoperative diagnosis were not predictive.
Conclusions: Hips implanted with an uncemented femoral stem, which has a uniquely exaggerated proximal taper angle, had an increased risk of both early and late postoperative periprosthetic femur fracture. The majority of patients with a fracture underwent reoperation or stem revision. The unique proximal geometry, lack of axial support from the smooth cylindrical distal stem as well as resorption of the hydroxyapatite coating and poor ongrowth with subsequent subsidence may contribute to increased risk of fracture. Although this particular stem has recently been discontinued by the manufacturer, these findings are important in regard to followup care for patients with this stem implanted as well as for future cementless stem design in general.
Level of evidence: Level III, therapeutic study.
Figures



Comment in
-
CORR Insights(®): Increased Risk of Periprosthetic Femur Fractures Associated With a Unique Cementless Stem Design.Clin Orthop Relat Res. 2015 Jun;473(6):2054-5. doi: 10.1007/s11999-014-4119-3. Epub 2015 Jan 6. Clin Orthop Relat Res. 2015. PMID: 25560961 Free PMC article. No abstract available.
Similar articles
-
High Survivorship With Cementless Stems and Cortical Strut Allografts for Large Femoral Bone Defects in Revision THA.Clin Orthop Relat Res. 2015 Sep;473(9):2990-3000. doi: 10.1007/s11999-015-4358-y. Epub 2015 May 27. Clin Orthop Relat Res. 2015. PMID: 26013152 Free PMC article.
-
Do Rerevision Rates Differ After First-time Revision of Primary THA With a Cemented and Cementless Femoral Component?Clin Orthop Relat Res. 2015 Nov;473(11):3391-8. doi: 10.1007/s11999-015-4245-6. Clin Orthop Relat Res. 2015. PMID: 25762017 Free PMC article.
-
Is There a Benefit to Modularity in 'Simpler' Femoral Revisions?Clin Orthop Relat Res. 2016 Feb;474(2):415-20. doi: 10.1007/s11999-015-4474-8. Clin Orthop Relat Res. 2016. PMID: 26245164 Free PMC article.
-
Management of periprosthetic femoral fractures following total hip arthroplasty: a review.Int Orthop. 2015 Oct;39(10):2005-10. doi: 10.1007/s00264-015-2979-0. Epub 2015 Aug 29. Int Orthop. 2015. PMID: 26318883 Review.
-
Undetected Intraoperative Periprosthetic Femoral Fractures in Patients Undergoing Primary Total Hip Arthroplasty: A Retrospective Case Series and Literature Review.Orthop Surg. 2023 Mar;15(3):758-765. doi: 10.1111/os.13646. Epub 2023 Jan 17. Orthop Surg. 2023. PMID: 36647808 Free PMC article. Review.
Cited by
-
Cementless tapered wedge stems in patients undergoing primary total hip arthroplasty with Dorr C bone-are complication risks increased?Ann Transl Med. 2023 Aug 30;11(10):349. doi: 10.21037/atm.2019.08.124. Epub 2019 Sep 23. Ann Transl Med. 2023. PMID: 37675300 Free PMC article.
-
Treatment of Vancouver type B2 periprosthetic femoral fractures.Int Orthop. 2015 Oct;39(10):1989-93. doi: 10.1007/s00264-015-2959-4. Epub 2015 Aug 20. Int Orthop. 2015. PMID: 26289168
-
Periprosthetic fractures: epidemiology and current treatment.Clin Cases Miner Bone Metab. 2017 May-Aug;14(2):189-196. doi: 10.11138/ccmbm/2017.14.1.189. Epub 2017 Oct 25. Clin Cases Miner Bone Metab. 2017. PMID: 29263732 Free PMC article. Review.
-
The learning curve following adoption of a novel short-stem prosthesis in total hip arthroplasty: implications on short-term patient outcomes.Eur J Orthop Surg Traumatol. 2019 May;29(4):819-825. doi: 10.1007/s00590-018-2355-z. Epub 2018 Dec 6. Eur J Orthop Surg Traumatol. 2019. PMID: 30523465
-
A comparative study about the incidence of dislocation and peri-prosthetic fracture between dual mobility versus standard cups after primary total hip arthroplasty.Int Orthop. 2019 Dec;43(12):2691-2695. doi: 10.1007/s00264-018-4279-y. Epub 2019 Jan 5. Int Orthop. 2019. PMID: 30612171
References
-
- Callaghan JJ, Dysart SH, Savory CG. The uncemented porous-coated anatomic total hip prosthesis. Two-year results of a prospective consecutive series. J Bone Joint Surg Am. 1988;70:337–346. - PubMed
-