Comparing velocity and fluid shear stress in a stenotic phantom with steady flow: phase-contrast MRI, particle image velocimetry and computational fluid dynamics
- PMID: 25502616
- DOI: 10.1007/s10334-014-0476-x
Comparing velocity and fluid shear stress in a stenotic phantom with steady flow: phase-contrast MRI, particle image velocimetry and computational fluid dynamics
Abstract
Object: This study aims to validate phase-contrast magnetic resonance imaging (PC-MRI) measurements of a steady flow through a severe stenotic phantom using particle image velocimetry (PIV) and computational fluid dynamics (CFD).
Materials and methods: The study was performed in an axisymmetric 87 % area stenosis model using an inlet Reynolds number (Re) of 160, corresponding to a jet Re of 444. Velocity patterns and estimated fluid shear stresses from three modalities were analyzed and compared qualitatively and quantitatively.
Results: Visual analysis via contour subtraction and Bland-Altman plots showed good agreement for flow velocities and less agreement for maximum shear stress (MSS). The Pearson's coefficients of correlation between PC-MRI and PIV were 0.97 for the velocity field and 0.82 for the MSS. The corresponding parameters between PC-MRI and CFD were 0.96 and 0.84, respectively.
Conclusion: Findings indicate that PC-MRI can be implemented to estimate velocity flow fields and MSS; however, this method is not sufficiently accurate to quantify the MSS at regions of high shear rate.
Similar articles
-
In vitro validation of flow measurement with phase contrast MRI at 3 tesla using stereoscopic particle image velocimetry and stereoscopic particle image velocimetry-based computational fluid dynamics.J Magn Reson Imaging. 2014 Jun;39(6):1477-85. doi: 10.1002/jmri.24322. Epub 2013 Oct 7. J Magn Reson Imaging. 2014. PMID: 24123721
-
Complex flow patterns in a real-size intracranial aneurysm phantom: phase contrast MRI compared with particle image velocimetry and computational fluid dynamics.NMR Biomed. 2012 Jan;25(1):14-26. doi: 10.1002/nbm.1706. Epub 2011 Apr 8. NMR Biomed. 2012. PMID: 21480417
-
Wall shear stress estimated with phase contrast MRI in an in vitro and in vivo intracranial aneurysm.J Magn Reson Imaging. 2013 Oct;38(4):876-84. doi: 10.1002/jmri.24051. Epub 2013 Feb 15. J Magn Reson Imaging. 2013. PMID: 23417769
-
Advanced blood flow assessment in Zebrafish via experimental digital particle image velocimetry and computational fluid dynamics modeling.Micron. 2020 Mar;130:102801. doi: 10.1016/j.micron.2019.102801. Epub 2019 Dec 14. Micron. 2020. PMID: 31864139 Review.
-
A Review of Arterial Phantom Fabrication Methods for Flow Measurement Using PIV Techniques.Ann Biomed Eng. 2018 Nov;46(11):1697-1721. doi: 10.1007/s10439-018-2085-8. Epub 2018 Jul 9. Ann Biomed Eng. 2018. PMID: 29987543 Review.
Cited by
-
Validation and Diagnostic Performance of a CFD-Based Non-invasive Method for the Diagnosis of Aortic Coarctation.Front Neuroinform. 2020 Dec 9;14:613666. doi: 10.3389/fninf.2020.613666. eCollection 2020. Front Neuroinform. 2020. PMID: 33362500 Free PMC article.
-
Hemodynamic analysis of carotid artery after endarterectomy: a preliminary and quantitative imaging study based on computational fluid dynamics and magnetic resonance angiography.Quant Imaging Med Surg. 2018 May;8(4):399-409. doi: 10.21037/qims.2018.05.02. Quant Imaging Med Surg. 2018. PMID: 29928605 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous