Accuracy of Intra-arterial and Subcutaneous Continuous Glucose Monitoring in Postoperative Cardiac Surgery Patients in the ICU
- PMID: 25539652
- PMCID: PMC4604543
- DOI: 10.1177/1932296814564993
Accuracy of Intra-arterial and Subcutaneous Continuous Glucose Monitoring in Postoperative Cardiac Surgery Patients in the ICU
Abstract
Background: The GluCath® intra-arterial continuous glucose monitoring (IA-CGM) system uses a novel quenched chemical fluorescence sensing mechanism to optically measure blood glucose when deployed in the radial artery. The aim of this study was to compare the accuracy of the IA-CGM and the FreeStyle Navigator® subcutaneous continuous glucose monitoring (SC-CGM) system with standard care.
Methods: After admission to the intensive care unit (ICU), the IA-CGM was inserted via a 20 gauge radial arterial study catheter and the SC-CGM was placed at the abdominal wall of postoperative cardiac surgery patients with an expected ICU length of stay > 24 hours. Each device was calibrated according to manufacturer instructions. Glucose values of both CGM systems were blinded for the clinical staff. Reference blood glucose samples were collected from the study catheter every 1-2 hours for at least 24 hours and analyzed on a Radiometer ABL blood gas analyzer.
Results: The IA-CGM and SC-CGM sensors were successfully inserted in 8 subjects. Accuracy assessment was performed with 183 paired points: 85.8% of the IA-CGM measurements and 84.2% of the SC-CGM measurements met ISO 15197:2003 glucometer criteria (within 20%) across a 79-248 mg/dl (4.4-13.8 mmol/L) glucose range. Overall ± SD mean absolute relative difference was 12.3 ± 11.3% for IA-CGM and 11.1 ± 8.3% for SC-CGM (difference -1.2%, 95% CI -3.3 to 0.8; P = .24).
Conclusions: The IA-CGM system directly measured arterial blood glucose and did not interfere with clinical care. However, accuracy was similar to that of the less invasive SC-CGM device.
Keywords: accuracy; continuous monitoring; glucose; intra-arterial; subcutaneous; vascular.
© 2014 Diabetes Technology Society.
Conflict of interest statement
Figures
References
-
- Marik PE, Preiser JC. Toward understanding tight glycemic control in the ICU: a systematic review and metaanalysis. Chest. 2010;137:544-551. - PubMed
-
- Brunner R, Kitzberger R, Miehsler W, Herkner H, Madl C, Holzinger U. Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients. Crit Care Med. 2011;39:659-664. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous