EZ-IO(®) intraosseous device implementation in German Helicopter Emergency Medical Service
- PMID: 25553609
- DOI: 10.1016/j.resuscitation.2014.12.015
EZ-IO(®) intraosseous device implementation in German Helicopter Emergency Medical Service
Abstract
Background: Intraosseous access (IO) is a rapid and safe alternative when peripheral venous access is difficult. Our aim was to summarize the first three years experience with the use of a semi-automatic IO device (EZ-IO(®)) in German Helicopter Emergency Medical Service (HEMS).
Methods: Included were all patients during study period (January 2009-December 2011) requiring an IO access performed by HEMS team. Outcome variables were IO rate, IO insertion success rates, site of IO access, type of EZ-IO(®) needle set used, strategy of vascular access, procedure related problems and operator's satisfaction.
Results: IO rate was 0.3% (348/120.923). Overall success rate was 99.6% with a first attempt success rate of 85.9%; there was only one failure (0.4%). There were three insertion sites: proximal tibia (87.2%), distal tibia (7.5%) and proximal humerus (5.3%). Within total study group IO was predominantly the second-line strategy (39% vs. 61%, p<0.001), but in children<7 years, in trauma cases and in cardiac arrest IO was more often first-line strategy (64% vs. 28%, p<0.001; 48% vs. 34%, p<0.032; 50% vs. 29%, p<0.002 respectively). Patients with IO access were significantly younger (41.7±28.7 vs. 56.5±24.4 years; p<0.001), more often male (63.2% vs. 57.7%; p=0.037), included more trauma cases (37.3% vs. 30.0%; p=0.003) and more often patients with a NACA-Score≥5 rating (77.0% vs. 18.6%; p<0.001). Patients who required IO access generally presented with more severely compromised vital signs associated with the need for more invasive resuscitation actions such as intubation, chest drains, CPR and defibrillation. In 93% EZ-IO(®) needle set handling was rated "good". Problems were reported in 1.6% (needle dislocation 0.8%, needle bending 0.4% and parafusion 0.4%).
Conclusions: The IO route was generally used in the most critically ill of patients. Our relatively low rate of usage would indicate that this would be compatible with the recommendations of established guidelines. The EZ-IO(®) intraosseous device proved feasible with a high success rate in adult and pediatric emergency patients in HEMS.
Keywords: EZ-IO(®); Emergency medicine; Helicopter Emergency Medical Service; Intraosseous vascular access; Vascular access.
Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Comparison of the Fluid Resuscitation Rate with and without External Pressure Using Two Intraosseous Infusion Systems for Adult Emergencies, the CITRIN (Comparison of InTRaosseous infusion systems in emergency medicINe)-Study.PLoS One. 2015 Dec 2;10(12):e0143726. doi: 10.1371/journal.pone.0143726. eCollection 2015. PLoS One. 2015. PMID: 26630579 Free PMC article.
-
EZ-IO(®) intraosseous device implementation in a pre-hospital emergency service: A prospective study and review of the literature.Resuscitation. 2013 Apr;84(4):440-5. doi: 10.1016/j.resuscitation.2012.11.006. Epub 2012 Nov 14. Resuscitation. 2013. PMID: 23160104 Review.
-
Intraosseous access EZ-IO in a prehospital emergency service.J Emerg Nurs. 2013 Sep;39(5):511-4. doi: 10.1016/j.jen.2012.03.005. Epub 2012 Oct 23. J Emerg Nurs. 2013. PMID: 23099011
-
Emergency intraosseous access in a helicopter emergency medical service: a retrospective study.Scand J Trauma Resusc Emerg Med. 2010 Oct 7;18:52. doi: 10.1186/1757-7241-18-52. Scand J Trauma Resusc Emerg Med. 2010. PMID: 20929544 Free PMC article.
-
Tibial Intraosseous Insertion in Pediatric Emergency Care: A Review Based upon Postmortem Computed Tomography.Prehosp Emerg Care. 2020 Sep-Oct;24(5):665-671. doi: 10.1080/10903127.2019.1698682. Epub 2020 Jan 7. Prehosp Emerg Care. 2020. PMID: 31774707 Review.
Cited by
-
[Current recommendations for basic/advanced life support : Addressing unanswered questions and future prospects].Med Klin Intensivmed Notfmed. 2016 Nov;111(8):670-681. doi: 10.1007/s00063-016-0216-2. Epub 2016 Sep 27. Med Klin Intensivmed Notfmed. 2016. PMID: 27677758 German.
-
Factors influencing the success and complications of intraosseous access in pediatric patients-a prospective nationwide surveillance study in Germany.Front Pediatr. 2023 Nov 29;11:1294322. doi: 10.3389/fped.2023.1294322. eCollection 2023. Front Pediatr. 2023. PMID: 38094190 Free PMC article.
-
Efficacy of intraosseous access for trauma resuscitation: a systematic review and meta-analysis.World J Emerg Surg. 2023 Mar 14;18(1):17. doi: 10.1186/s13017-023-00487-7. World J Emerg Surg. 2023. PMID: 36918947 Free PMC article.
-
Intraosseous access in the resuscitation of trauma patients: a literature review.Eur J Trauma Emerg Surg. 2021 Feb;47(1):47-55. doi: 10.1007/s00068-020-01327-y. Epub 2020 Feb 20. Eur J Trauma Emerg Surg. 2021. PMID: 32078703 Review.
-
Comparison of the Fluid Resuscitation Rate with and without External Pressure Using Two Intraosseous Infusion Systems for Adult Emergencies, the CITRIN (Comparison of InTRaosseous infusion systems in emergency medicINe)-Study.PLoS One. 2015 Dec 2;10(12):e0143726. doi: 10.1371/journal.pone.0143726. eCollection 2015. PLoS One. 2015. PMID: 26630579 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical