Fluidics comparison between dual pneumatic and spring return high-speed vitrectomy systems
- PMID: 25559512
- DOI: 10.3928/23258160-20150101-11
Fluidics comparison between dual pneumatic and spring return high-speed vitrectomy systems
Abstract
Background and objective: To compare the water and vitreous flow rates and duty cycle (DC) between two ultrahigh-speed vitrectomy systems: pneumatic with spring return (SR) and dual pneumatic (DP) probes.
Materials and methods: The flow rate was calculated using a high-sampling precision balance that measured the mass of water and vitreous removed from a vial by a vitreous cutter. Frame-by-frame analysis of a high-speed video of the cutter was used to determine the DC. Three cutters of each gauge (20, 23, and 25 G) were tested with an SR and a DP system using the standard DC setting (biased open) at 0 (water only), 1,000, 2,000, 3,000, 4,000, and 5,000 cuts per minute (CPM) with aspiration levels of 100, 200, 300, 400, 500, and 600 mm Hg.
Results: The DC was slightly higher with the SR system using most parameters and gauges although without statistical significance. The water flow rate was somewhat higher with the SR system, except for 25 G with 4,000 and 5,000 CPM. The vitreous flow rate was similar using most parameters, with the SR system showing higher flows at lower cut rates (1,000-3,000 CPM).
Conclusions: SR and DP systems produced similar water and vitreous flow rates. Additional studies in human eyes are necessary to confirm these findings.
Copyright 2015, SLACK Incorporated.
Similar articles
-
Performance analysis of millennium vitreous enhancer™ system.Ophthalmic Surg Lasers Imaging. 2011 Mar-Apr;42(2):162-7. doi: 10.3928/15428877-20101223-03. Epub 2010 Dec 30. Ophthalmic Surg Lasers Imaging. 2011. PMID: 21210579
-
Performance analysis of ultrahigh-speed vitreous cutter system.Retina. 2013 May;33(5):928-32. doi: 10.1097/IAE.0b013e31826f069e. Retina. 2013. PMID: 23416511
-
Performance analysis of new-generation vitreous cutters.Graefes Arch Clin Exp Ophthalmol. 2008 Jan;246(1):61-7. doi: 10.1007/s00417-007-0672-8. Epub 2007 Sep 18. Graefes Arch Clin Exp Ophthalmol. 2008. PMID: 17876598
-
Vitrectomy fluidics.Ophthalmologica. 2011;226 Suppl 1:27-35. doi: 10.1159/000328207. Epub 2011 Jul 22. Ophthalmologica. 2011. PMID: 21778777 Review.
-
[Pars plana vitrectomy-from suction cutting systems to ultrasound technology : Vitesse-a new form of vitrectomy based on ultrasound technology].Ophthalmologe. 2021 Jul;118(7):741-746. doi: 10.1007/s00347-021-01377-6. Epub 2021 Apr 20. Ophthalmologe. 2021. PMID: 33877403 Free PMC article. Review. German.
Cited by
-
Fluidics of Single and Double Blade Guillotine Vitrectomy Probes in Balanced Salt Solution and Artificial Vitreous.Transl Vis Sci Technol. 2018 Dec 6;7(6):19. doi: 10.1167/tvst.7.6.19. eCollection 2018 Nov. Transl Vis Sci Technol. 2018. PMID: 30564508 Free PMC article.
-
Vitreoretinal instruments: vitrectomy cutters, endoillumination and wide-angle viewing systems.Int J Retina Vitreous. 2016 Dec 5;2:28. doi: 10.1186/s40942-016-0052-9. eCollection 2016. Int J Retina Vitreous. 2016. PMID: 27980854 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials