A mitral annulus tracking approach for navigation of off-pump beating heart mitral valve repair
- PMID: 25563285
- DOI: 10.1118/1.4904022
A mitral annulus tracking approach for navigation of off-pump beating heart mitral valve repair
Abstract
Purpose: To develop and validate a real-time mitral valve annulus (MVA) tracking approach based on biplane transesophageal echocardiogram (TEE) data and magnetic tracking systems (MTS) to be used in minimally invasive off-pump beating heart mitral valve repair (MVR).
Methods: The authors' guidance system consists of three major components: TEE, magnetic tracking system, and an image guidance software platform. TEE provides real-time intraoperative images to show the cardiac motion and intracardiac surgical tools. The magnetic tracking system tracks the TEE probe and the surgical tools. The software platform integrates the TEE image planes and the virtual model of the tools and the MVA model on the screen. The authors' MVA tracking approach, which aims to update the MVA model in near real-time, comprises of three steps: image based gating, predictive reinitialization, and registration based MVA tracking. The image based gating step uses a small patch centered at each MVA point in the TEE images to identify images at optimal cardiac phases for updating the position of the MVA. The predictive reinitialization step uses the position and orientation of the TEE probe provided by the magnetic tracking system to predict the position of the MVA points in the TEE images and uses them for the initialization of the registration component. The registration based MVA tracking step aims to locate the MVA points in the images selected by the image based gating component by performing image based registration.
Results: The validation of the MVA tracking approach was performed in a phantom study and a retrospective study on porcine data. In the phantom study, controlled translations were applied to the phantom and the tracked MVA was compared to its "true" position estimated based on a magnetic sensor attached to the phantom. The MVA tracking accuracy was 1.29 ± 0.58 mm when the translation distance is about 1 cm, and increased to 2.85 ± 1.19 mm when the translation distance is about 3 cm. In the study on porcine data, the authors compared the tracked MVA to a manually segmented MVA. The overall accuracy is 2.37 ± 1.67 mm for single plane images and 2.35 ± 1.55 mm for biplane images. The interoperator variation in manual segmentation was 2.32 ± 1.24 mm for single plane images and 1.73 ± 1.18 mm for biplane images. The computational efficiency of the algorithm on a desktop computer with an Intel(®) Xeon(®) CPU @3.47 GHz and an NVIDIA GeForce 690 graphic card is such that the time required for registering four MVA points was about 60 ms.
Conclusions: The authors developed a rapid MVA tracking algorithm for use in the guidance of off-pump beating heart transapical mitral valve repair. This approach uses 2D biplane TEE images and was tested on a dynamic heart phantom and interventional porcine image data. Results regarding the accuracy and efficiency of the authors' MVA tracking algorithm are promising, and fulfill the requirements for surgical navigation.
Similar articles
-
Augmented reality image guidance improves navigation for beating heart mitral valve repair.Innovations (Phila). 2012 Jul-Aug;7(4):274-81. doi: 10.1097/IMI.0b013e31827439ea. Innovations (Phila). 2012. PMID: 23123995
-
Ultrasound image and augmented reality guidance for off-pump, closed, beating, intracardiac surgery.Artif Organs. 2008 Nov;32(11):840-5. doi: 10.1111/j.1525-1594.2008.00639.x. Artif Organs. 2008. PMID: 18959675
-
Ultrasound guidance for beating heart mitral valve repair augmented by synthetic dynamic CT.IEEE Trans Med Imaging. 2015 Oct;34(10):2025-35. doi: 10.1109/TMI.2015.2412465. Epub 2015 Mar 11. IEEE Trans Med Imaging. 2015. PMID: 25775487
-
Intraoperative transesophageal echocardiography for surgical repair of mitral regurgitation.J Am Soc Echocardiogr. 2014 Apr;27(4):345-66. doi: 10.1016/j.echo.2014.01.005. Epub 2014 Feb 15. J Am Soc Echocardiogr. 2014. PMID: 24534653 Review.
-
Multiplane transesophageal echocardiographic evaluation of mitral valve disease.Am J Card Imaging. 1995 Apr;9(2):121-8. Am J Card Imaging. 1995. PMID: 7795376 Review.
Cited by
-
Navigation guidance for ventricular septal defect closure in heart phantoms.Int J Comput Assist Radiol Surg. 2022 Oct;17(10):1947-1956. doi: 10.1007/s11548-022-02711-2. Epub 2022 Jul 7. Int J Comput Assist Radiol Surg. 2022. PMID: 35798998
-
An Overview on Image Registration Techniques for Cardiac Diagnosis and Treatment.Cardiol Res Pract. 2018 Aug 8;2018:1437125. doi: 10.1155/2018/1437125. eCollection 2018. Cardiol Res Pract. 2018. PMID: 30159169 Free PMC article. Review.
-
3D localization of vena contracta using Doppler ICE imaging in tricuspid valve interventions.Int J Comput Assist Radiol Surg. 2022 Sep;17(9):1569-1577. doi: 10.1007/s11548-022-02660-w. Epub 2022 May 19. Int J Comput Assist Radiol Surg. 2022. PMID: 35588338 Free PMC article.
-
Patient-specific cardiac phantom for clinical training and preprocedure surgical planning.J Med Imaging (Bellingham). 2018 Apr;5(2):021222. doi: 10.1117/1.JMI.5.2.021222. Epub 2018 Mar 23. J Med Imaging (Bellingham). 2018. PMID: 29594183 Free PMC article.
-
Novel Intracardiac Ultrasound Images Developed on a Cardiac Ultrasound Simulator and Validated in Live Horses.J Vet Intern Med. 2025 May-Jun;39(3):e70087. doi: 10.1111/jvim.70087. J Vet Intern Med. 2025. PMID: 40269662 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources