Motion correction strategies for integrated PET/MR
- PMID: 25572092
- DOI: 10.2967/jnumed.114.146787
Motion correction strategies for integrated PET/MR
Abstract
Integrated whole-body PET/MR facilitates the implementation of a broad variety of respiratory motion correction strategies, taking advantage of the strengths of both modalities. The goal of this study was the quantitative evaluation with clinical data of different MR- and PET-data-based motion correction strategies for integrated PET/MR.
Methods: The PET and MR data of 20 patients were simultaneously acquired for 10 min on an integrated PET/MR system after administration of (18)F-FDG or (68)Ga-DOTANOC. Respiratory traces recorded with a bellows were compared against MR self-gating signals and signals extracted from PET raw data with the sensitivity method, by applying principal component analysis (PCA) or Laplacian eigenmaps and by using a novel variation combining the former and either of the latter two. Gated sinograms and MR images were generated accordingly, followed by image registration to derive MR motion models. Corrected PET images were reconstructed by incorporating this information into the reconstruction. An optical flow algorithm was applied for PET-based motion correction. Gating and motion correction were evaluated by quantitative analysis of apparent tracer uptake, lesion volume, displacement, contrast, and signal-to-noise ratio.
Results: The correlation between bellows- and MR-based signals was 0.63 ± 0.19, and that between MR and the sensitivity method was 0.52 ± 0.26. Depending on the PET raw-data compression, the average correlation between MR and PCA ranged from 0.25 ± 0.30 to 0.58 ± 0.33, and the range was 0.25 ± 0.30 to 0.42 ± 0.34 if Laplacian eigenmaps were applied. By combining the sensitivity method and PCA or Laplacian eigenmaps, the maximum average correlation to MR could be increased to 0.74 ± 0.21 and 0.70 ± 0.19, respectively. The selection of the best PET-based signal for each patient yielded an average correlation of 0.80 ± 0.13 with MR. Using the best PET-based respiratory signal for gating, mean tracer uptake increased by 17 ± 19% for gating, 13 ± 10% for MR-based motion correction, and 18 ± 15% for PET-based motion correction, compared with the static images. Lesion volumes were 76 ± 31%, 83 ± 18%, and 74 ± 22% of the sizes in the static images for gating, MR-based motion correction, and PET-based motion correction, respectively.
Conclusion: Respiratory traces extracted from MR and PET data are comparable to those based on external sensors. The proposed PET-driven gating method improved respiratory signals and overall stability. Consistent results from MR- and PET-based correction methods enable more flexible PET/MR scan protocols while achieving higher PET image quality.
Keywords: PET/MR; motion correction; multimodal imaging; respiratory gating.
© 2015 by the Society of Nuclear Medicine and Molecular Imaging, Inc.
Similar articles
-
Practical PET Respiratory Motion Correction in Clinical PET/MR.J Nucl Med. 2015 Jun;56(6):890-6. doi: 10.2967/jnumed.114.151779. Epub 2015 May 7. J Nucl Med. 2015. PMID: 25952740
-
Reconstruction-Incorporated Respiratory Motion Correction in Clinical Simultaneous PET/MR Imaging for Oncology Applications.J Nucl Med. 2015 Jun;56(6):884-9. doi: 10.2967/jnumed.114.153007. Epub 2015 Apr 23. J Nucl Med. 2015. PMID: 25908830
-
Respiratory motion correction in oncologic PET using T1-weighted MR imaging on a simultaneous whole-body PET/MR system.J Nucl Med. 2013 Mar;54(3):464-71. doi: 10.2967/jnumed.112.105296. Epub 2013 Jan 3. J Nucl Med. 2013. PMID: 23287577
-
Motion correction using anatomical information in PET/CT and PET/MR hybrid imaging.Q J Nucl Med Mol Imaging. 2016 Mar;60(1):12-24. Epub 2015 Nov 17. Q J Nucl Med Mol Imaging. 2016. PMID: 26576736 Review.
-
MR-Based Cardiac and Respiratory Motion-Compensation Techniques for PET-MR Imaging.PET Clin. 2016 Apr;11(2):179-91. doi: 10.1016/j.cpet.2015.09.004. Epub 2016 Jan 26. PET Clin. 2016. PMID: 26952730 Review.
Cited by
-
Evaluating different methods of MR-based motion correction in simultaneous PET/MR using a head phantom moved by a robotic system.EJNMMI Phys. 2022 Mar 3;9(1):15. doi: 10.1186/s40658-022-00442-6. EJNMMI Phys. 2022. PMID: 35239047 Free PMC article.
-
Motion-corrected simultaneous cardiac positron emission tomography and coronary MR angiography with high acquisition efficiency.Magn Reson Med. 2018 Jan;79(1):339-350. doi: 10.1002/mrm.26690. Epub 2017 Apr 20. Magn Reson Med. 2018. PMID: 28426162 Free PMC article.
-
Advances in PET/MR instrumentation and image reconstruction.Br J Radiol. 2018 Jan;91(1081):20160363. doi: 10.1259/bjr.20160363. Epub 2016 Jul 22. Br J Radiol. 2018. PMID: 27376170 Free PMC article. Review.
-
Motion correction in simultaneous PET/MR brain imaging using sparsely sampled MR navigators: a clinically feasible tool.EJNMMI Phys. 2015 Dec;2(1):14. doi: 10.1186/s40658-015-0118-z. Epub 2015 Jul 16. EJNMMI Phys. 2015. PMID: 26501815 Free PMC article.
-
Cardiac and Respiratory Motion Correction for Simultaneous Cardiac PET/MR.J Nucl Med. 2017 May;58(5):846-852. doi: 10.2967/jnumed.115.171728. Epub 2017 Feb 9. J Nucl Med. 2017. PMID: 28183991 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical