Abstract
Purpose: To compare flap morphology created by the WaveLight FS200 femtosecond laser and the VisuMax femtosecond laser, assessing the uniformity, accuracy, and predictability of flap creation.
Methods: A total of 400 eyes had corneal flaps created with the WaveLight FS200 femtosecond laser (200 eyes) or the VisuMax femtosecond laser (200 eyes). The desired flap thickness was 110 μm. At 1 week postoperatively, all eyes were evaluated with RTVue Fourier-domain optical coherence tomography. Dimensions of the flaps were tested for their regularity, uniformity, accuracy, and predictability comparison.
Results: One week after surgery, the central flap thickness and the mean flap thickness of the FS200 group were 105.4 ± 3.4 μm and 105.7 ± 2.6 μm, respectively. They were both thinner than those of the VisuMax group, which were 110.8 ± 3.9 μm and 111.3 ± 2.3 μm, respectively. The mean deviation between the achieved and attempted flap thickness of the FS200 group (5.2 ± 1.9 μm) was greater than that of the VisuMax group (3.2 ± 1.8 μm). Flap thickness measurements at 36 points in both groups were close to the intended thickness. Morphology of the flaps in the 0-, 45-, 90-, and 135-degree lines created by the FS200 femtosecond laser and VisuMax femtosecond laser were uniform and regular.
Conclusions: Flap dimensions created by the WaveLight FS200 femtosecond laser and VisuMax femtosecond laser were uniform and regular. Although the flap thickness created by the FS200 was less than that created by the VisuMax, measurements of both femtosecond lasers were close to the intended thickness.
Similar articles
-
Comparison of corneal flap morphology using AS-OCT in LASIK with the WaveLight FS200 femtosecond laser versus a mechanical microkeratome.J Refract Surg. 2013 May;29(5):320-4. doi: 10.3928/1081597X-20130415-03. J Refract Surg. 2013. PMID: 23659230
-
Predictability of corneal flap thickness in laser in situ keratomileusis using a 200 kHz femtosecond laser.J Cataract Refract Surg. 2013 Mar;39(3):378-85. doi: 10.1016/j.jcrs.2012.10.041. Epub 2013 Jan 23. J Cataract Refract Surg. 2013. PMID: 23352500
-
Prospective contralateral eye study to compare 80- and 120-μm flap LASIK using the VisuMax femtosecond laser.J Refract Surg. 2013 Jul;29(7):462-8. doi: 10.3928/1081597X-20130617-04. J Refract Surg. 2013. PMID: 23820228 Clinical Trial.
-
Complications of femtosecond-assisted laser in-situ keratomileusis flaps.Semin Ophthalmol. 2014 Sep-Nov;29(5-6):363-75. doi: 10.3109/08820538.2014.959194. Semin Ophthalmol. 2014. PMID: 25325862 Review.
-
Microkeratome versus femtosecond flaps: accuracy and complications.Curr Opin Ophthalmol. 2014 Jul;25(4):270-4. doi: 10.1097/ICU.0000000000000070. Curr Opin Ophthalmol. 2014. PMID: 24837579 Review.
Cited by
-
Combined Phototherapeutic Keratectomy and Peripheral Anterior Stromal Puncture for the Treatment of Recurrent Corneal Erosion Syndrome.Korean J Ophthalmol. 2020 Aug;34(4):297-303. doi: 10.3341/kjo.2020.0023. Korean J Ophthalmol. 2020. PMID: 32783422 Free PMC article.
-
Safety and Precision of Two Different Flap-morphologies Created During Low Energy Femtosecond Laser-assisted LASIK.J Ophthalmic Vis Res. 2023 Feb 21;18(1):3-14. doi: 10.18502/jovr.v18i1.12720. eCollection 2023 Jan-Mar. J Ophthalmic Vis Res. 2023. PMID: 36937201 Free PMC article.
-
Flap Thickness and the Risk of Complications in Mechanical Microkeratome and Femtosecond Laser In Situ Keratomileusis: A Literature Review and Statistical Analysis.Diagnostics (Basel). 2021 Aug 31;11(9):1588. doi: 10.3390/diagnostics11091588. Diagnostics (Basel). 2021. PMID: 34573930 Free PMC article. Review.
-
The evolution of corneal and refractive surgery with the femtosecond laser.Eye Vis (Lond). 2015 Jul 14;2:12. doi: 10.1186/s40662-015-0022-6. eCollection 2015. Eye Vis (Lond). 2015. PMID: 26605365 Free PMC article. Review.
-
Comparison of Laser In Situ Keratomileusis Flap Morphology and Predictability by WaveLight FS200 Femtosecond Laser and Moria Microkeratome: An Anterior Segment Optical Coherence Tomography Study.Korean J Ophthalmol. 2019 Apr;33(2):113-121. doi: 10.3341/kjo.2018.0035. Korean J Ophthalmol. 2019. PMID: 30977320 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources