Patient's Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections
- PMID: 25610557
- PMCID: PMC4291604
- DOI: 10.15171/jcvtr.2014.020
Patient's Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections
Abstract
Introduction: We aimed to determine angiography projections with lower Dose Area Product (DAP) rate by measuring the mean DAP and fluoroscopy times in coronary angiography (CAG) and percutaneous coronary intervention (PCI) and calculating DAP rate in different projections.
Methods: DAP and fluoroscopy times were measured in all employed projections in real-time in 75 patients who underwent CAG or PCI by a single cardiologist in Madani Cardiovascular University Hospital (45 in CAG group and 30 in PCI group). DAP rate was calculated in both groups and in all projections. The projections with highest and lowest DAP rate were determined.
Results: Mean DAP was 436.73±315.85 dGy×cm(2) in CAG group and 643.26±359.58 dGy×cm(2) in PCI group. The projection 40° LAO/0° had the highest DAP rate in CAG group (28.98 dGy×cm(2)/ sec) and it was highest in 20° RAO/30° CR in PCI group (29.83 dGy×cm(2)/sec). The latter projection was also the most employed projection in PCI group.
Conclusion: The amount of radiation dose in this study is in consistent with the previous reports. Specific angiographic projections expose patients to significantly higher radiation and they should be avoided and replaced by less irradiating projections whenever possible.
Keywords: Angiography; Angioplasty; Interventional; Radiation.
Similar articles
-
Effective and equivalent organ doses in patients undergoing coronary angiography and percutaneous coronary intervention.Med Phys. 2011 Apr;38(4):2168-75. doi: 10.1118/1.3561508. Med Phys. 2011. PMID: 21626950
-
Identification of less-irradiating tube angulations in invasive cardiology.J Am Coll Cardiol. 2004 Oct 6;44(7):1420-8. doi: 10.1016/j.jacc.2004.06.057. J Am Coll Cardiol. 2004. PMID: 15464322
-
Radiation exposure to patients undergoing percutaneous coronary interventions: are current reference values too high?Herz. 2004 Mar;29(2):208-17. doi: 10.1007/s00059-004-2552-x. Herz. 2004. PMID: 15054590
-
Patient doses and dosimetric evaluations in interventional cardiology.Phys Med. 2009 Mar;25(1):31-42. doi: 10.1016/j.ejmp.2008.03.002. Epub 2008 Apr 25. Phys Med. 2009. PMID: 18440265
-
Anthropometric parameters and radiation doses during percutaneous coronary procedures.Phys Med. 2022 Aug;100:164-175. doi: 10.1016/j.ejmp.2022.06.013. Epub 2022 Jul 25. Phys Med. 2022. PMID: 35901630
Cited by
-
Radiation doses and estimated risk from angiographic projections during coronary angiography performed using novel flat detector.J Appl Clin Med Phys. 2016 May 8;17(3):433-441. doi: 10.1120/jacmp.v17i3.5926. J Appl Clin Med Phys. 2016. PMID: 27167263 Free PMC article.
-
Comparison of the patient radiation exposure during coronary angiography and angioplasty procedures using trans-radial and trans-femoral access.J Cardiovasc Thorac Res. 2016;8(2):77-82. doi: 10.15171/jcvtr.2016.15. Epub 2016 Jun 28. J Cardiovasc Thorac Res. 2016. PMID: 27489601 Free PMC article.
-
Beam projections and radiation exposure in transradial and transfemoral approaches during coronary angiography.Anatol J Cardiol. 2017 Oct;18(4):298-303. doi: 10.14744/AnatolJCardiol.2017.7724. Epub 2017 Aug 2. Anatol J Cardiol. 2017. PMID: 28777096 Free PMC article.
References
-
- O’Gara PT, Kushner FG, Ascheim DD, Casey DE, Chung MK, de Lemos JA. et al. 2013 ACCF/AHA guideline for the management of ST-elevation myocardial infarction: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2013;61:e78–e140. doi: 10.1161/cir.0b013e3182742c84. - DOI - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous