Reliable and accurate CD4+ T cell count and percent by the portable flow cytometer CyFlow MiniPOC and "CD4 Easy Count Kit-Dry", as revealed by the comparison with the gold standard dual platform technology
- PMID: 25622041
- PMCID: PMC4306486
- DOI: 10.1371/journal.pone.0116848
Reliable and accurate CD4+ T cell count and percent by the portable flow cytometer CyFlow MiniPOC and "CD4 Easy Count Kit-Dry", as revealed by the comparison with the gold standard dual platform technology
Abstract
Background: An accurate and affordable CD4+ T cells count is an essential tool in the fight against HIV/AIDS. Flow cytometry (FCM) is the "gold standard" for counting such cells, but this technique is expensive and requires sophisticated equipment, temperature-sensitive monoclonal antibodies (mAbs) and trained personnel. The lack of access to technical support and quality assurance programs thus limits the use of FCM in resource-constrained countries. We have tested the accuracy, the precision and the carry-over contamination of Partec CyFlow MiniPOC, a portable and economically affordable flow cytometer designed for CD4+ count and percentage, used along with the "CD4% Count Kit-Dry".
Materials and methods: Venous blood from 59 adult HIV+ patients (age: 25-58 years; 43 males and 16 females) was collected and stained with the "MiniPOC CD4% Count Kit-Dry". CD4+ count and percentage were then determined in triplicate by the CyFlow MiniPOC. In parallel, CD4 count was performed using mAbs and a CyFlow Counter, or by a dual platform system (from Beckman Coulter) based upon Cytomic FC500 ("Cytostat tetrachrome kit" for mAbs) and Coulter HmX Hematology Analyzer (for absolute cell count).
Results: The accuracy of CyFlow MiniPOC against Cytomic FC500 showed a correlation coefficient (CC) of 0.98 and 0.97 for CD4+ count and percentage, respectively. The accuracy of CyFlow MiniPOC against CyFlow Counter showed a CC of 0.99 and 0.99 for CD4 T cell count and percentage, respectively. CyFlow MiniPOC showed an excellent repeatability: CD4+ cell count and percentage were analyzed on two instruments, with an intra-assay precision below ± 5% deviation. Finally, there was no carry-over contamination for samples at all CD4 values, regardless of their position in the sequence of analysis.
Conclusion: The cost-effective CyFlow MiniPOC produces rapid, reliable and accurate results that are fully comparable with those from highly expensive dual platform systems.
Conflict of interest statement
Figures








Similar articles
-
Laboratory and field evaluation of the Partec CyFlow miniPOC for absolute and relative CD4 T-cell enumeration.PLoS One. 2015 Feb 17;10(2):e0116663. doi: 10.1371/journal.pone.0116663. eCollection 2015. PLoS One. 2015. PMID: 25688553 Free PMC article.
-
Comparison of two alternative methods for CD4+ T-cell determination (Coulter manual CD4 count and CyFlow) against standard dual platform flow cytometry in Uganda.Cytometry B Clin Cytom. 2006 May;70(3):163-9. doi: 10.1002/cyto.b.20093. Cytometry B Clin Cytom. 2006. PMID: 16498672
-
Evaluation of a new single-parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1-infected Thai patients.Clin Diagn Lab Immunol. 2005 Dec;12(12):1416-24. doi: 10.1128/CDLI.12.12.1416-1424.2005. Clin Diagn Lab Immunol. 2005. PMID: 16339065 Free PMC article.
-
Thirty-five years of CD4 T-cell counting in HIV infection: From flow cytometry in the lab to point-of-care testing in the field.Cytometry B Clin Cytom. 2017 Nov;92(6):437-444. doi: 10.1002/cyto.b.21400. Epub 2016 Aug 3. Cytometry B Clin Cytom. 2017. PMID: 27406947 Review.
-
Evaluation of a flow cytometry method for CD4 T cell enumeration based on volumetric primary CD4 gating using thermoresistant reagents.J Immunol Methods. 2011 Sep 30;372(1-2):7-13. doi: 10.1016/j.jim.2011.07.012. Epub 2011 Jul 29. J Immunol Methods. 2011. PMID: 21835181 Review.
Cited by
-
Establishing Reference Values for Peripheral Blood Lymphocyte Subsets of Healthy Children in China Using a Single Platform.J Immunol Res. 2022 Aug 17;2022:5603566. doi: 10.1155/2022/5603566. eCollection 2022. J Immunol Res. 2022. PMID: 36033395 Free PMC article.
-
Nut and Bolt Microfluidics with Helical Minichannel for Counting CD4+ T-Cells.Bioengineering (Basel). 2019 Mar 15;6(1):24. doi: 10.3390/bioengineering6010024. Bioengineering (Basel). 2019. PMID: 30875995 Free PMC article.
-
An inexpensive, customizable microscopy system for the automated quantification and characterization of multiple adherent cell types.PeerJ. 2018 Jun 5;6:e4937. doi: 10.7717/peerj.4937. eCollection 2018. PeerJ. 2018. PMID: 29888136 Free PMC article.
-
The Parasol Protocol: An Implementation Science Study of HIV Continuum of Care Interventions for Gay Men and Transgender Women in Burma/Myanmar.JMIR Res Protoc. 2017 May 17;6(5):e90. doi: 10.2196/resprot.7642. JMIR Res Protoc. 2017. PMID: 28526661 Free PMC article.
-
Modulation of Tregs and iNKT by Fingolimod in Multiple Sclerosis Patients.Cells. 2021 Nov 26;10(12):3324. doi: 10.3390/cells10123324. Cells. 2021. PMID: 34943831 Free PMC article.
References
-
- UNAIDS (2013) UNAIDS report on the global AIDS epidemic 2013.
-
- World Health Organization (2010). Antiretroviral therapy for HIV infection in adults and adolescents: Recommendations for a public health approach: 2010 revision. Geneva, Switzerland: - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials