Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2015 Jan;120(1):66-75.
doi: 10.1213/ANE.0000000000000477.

Metrology in medicine: from measurements to decision, with specific reference to anesthesia and intensive care

Affiliations
Free PMC article
Review

Metrology in medicine: from measurements to decision, with specific reference to anesthesia and intensive care

Pierre Squara et al. Anesth Analg. 2015 Jan.
Free PMC article

Abstract

Metrology is the science of measurements. Although of critical importance in medicine and especially in critical care, frequent confusion in terms and definitions impact either interphysician communications or understanding of manufacturers' and engineers' instructions and limitations when using devices. In this review, we first list the terms defined by the International Bureau of Weights and Measures regarding quantities and units, measurements, devices for measurement, properties of measuring devices, and measurement standards. The traditional tools for assessing the most important measurement quality criteria are also reviewed with clinical examples for diagnosis, alarm, and titration purposes, as well as for assessing the uncertainty of reference methods.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflicts of interest.

Figures

Figure 1.
Figure 1.
Schematic representation of different types of systematic errors. Typically, these lines are these observed when there is an issue in calibration (blue line), in amplification (red line), or when insufficient sensing is corrected by adding a constant (orange line).
Figure 2.
Figure 2.
Schematic representation of 2 types of random errors. Blue plots, small random error (2σ = 0.6) typically seen when an independent noise is properly filtered. In red plots, the variability in proportion to the measurement value indicating that the random noise is getting into the signal. In these examples, both data sets have correlation slopes on the identity line (black dotted line), indicating no systematic errors (x-axis and y-axis with arbitrary units).
Figure 3.
Figure 3.
Schematic representation of the different types of measurement errors with indication of the formula by which it is derived and the corresponding quality criteria. m = mean; σ = standard deviation of replicate measurements on the same object. In this figure, the variability is expressed by ±2σ/m to fit with a confidence interval of 95%.
Figure 4.
Figure 4.
Clark error grid for insulin dosing. Except for 2 measurements in zones C and D each, agreement between the 2 methods is acceptable for diabetes care in this example. (Reproduced from Ref. 20, with permission.)
Figure 5.
Figure 5.
Bland-Altman plots. Mean difference = 17 ± 40 mg/dL. Limits of agreements (2σ) of the measurement error/ averaged value = 51%. (Reproduced from Ref. 20, with permission.)
Figure 6.
Figure 6.
Modified error grid for the same data as in Figure 4. This error grid shows that the test device would have resulted in unacceptable, serious protocol violation, or even life-threatening therapy decisions in 10 cases, which may be considered unacceptable for clinical use in the intensive care unit. (Reproduced from Ref. 20, with permission.)
Figure 7.
Figure 7.
Bland-Altman plot of mean arterial pressure measured by a noninvasive method (mBP) and the invasive method (mBPIA). Difference = −3.7 ± 6.2 mm Hg, ±10.3%.
Figure 8.
Figure 8.
Same data as in Figure 6. Overlaid Bland-Altman plots of repeated measurements for each measurement technique (blue mBP; red = mBPIA). mBP: CR = 6.53, mBPIA: CR = 15.65

References

    1. Bureau International des poiuds et Mesures; Joint Committee for Guides in Metrology. International Vocabulary of Metrology—Basic and General Concepts and Associated Terms (VIM).Paris: 2012.
    1. Wachtel RE, Dexter F. Difficulties and challenges associated with literature searches in operating room management, complete with recommendations. Anesth Analg. 2013;117:1460–79. - PubMed
    1. BIMP. International System of Units. 2013.
    1. Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD. Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009;35:1801–8. - PubMed
    1. Cox DJ, Clarke WL, Gonder-Frederick L, Pohl S, Hoover C, Snyder A, Zimbelman L, Carter WR, Bobbitt S, Pennebaker J. Accuracy of perceiving blood glucose in IDDM. Diabetes Care. 1985;8:529–36. - PubMed