Evaluation of different diameter arterial tubing and arterial cannulae in simulated neonatal/pediatric cardiopulmonary bypass circuits
- PMID: 25626579
- DOI: 10.1111/aor.12446
Evaluation of different diameter arterial tubing and arterial cannulae in simulated neonatal/pediatric cardiopulmonary bypass circuits
Abstract
The objective of this study is to evaluate three different diameters of arterial tubing and three diameters of arterial cannulae in terms of pressure drop, and hemodynamic energy delivery in simulated neonatal/pediatric cardiopulmonary bypass (CPB) circuits. The CPB circuit consisted of a Terumo Capiox Baby FX05 oxygenator (Terumo Corporation, Tokyo, Japan), arterial tubing (1/4 in, 3/16 in, or 1/8 in × 150 cm), and a Medtronic Bio-Medicus arterial cannula (8, 10, or 12 Fr; Medtronic, Inc., Minneapolis, MN, USA). The pseudo patient's pressure was maintained at 50 mm Hg. The circuit was primed using lactated Ringer's solution and heparinized packed human red blood cells (hematocrit 30%). Trials were conducted at different flow rates and temperatures (35 and 28°C). Flow and pressure data were collected using a custom-based data acquisition system. Using 8 Fr arterial cannula at 500 mL/min, small diameter arterial tubing generated higher circuit pressure (294.6 ± 0.1 mm Hg [1/8 in], 213.5 ± 0.5 mm Hg [3/16 in], 208.4 ± 0.4 mm Hg [1/4 in] at 35°C) and arterial line pressure drop (158.3 ± 0.1 mm Hg [1/8 in], 79.6 ± 0.1 mm Hg [3/16 in], 62.1 ± 0.1 mm Hg [1/4 in] at 35°C). Using 10 Fr arterial cannula at 1000 mL/min, pre-oxygenator pressures were 266.8 ± 0.2 mm Hg (3/16 in) and 248.0 ± 0.3 mm Hg (1/4 in); arterial line pressure drops were 111.6 ± 0.0 mm Hg (3/16 in) and 74.0 ± 0.1 mm Hg (1/4 in) at 35°C. When using 12 Fr arterial cannula at 1500 mL/min, preoxygenator pressures reached 324.4 ± 0.3 mm Hg (3/16 in) and 302.5 ± 0.4 mm Hg (1/4 in); arterial line pressure drops were 154.0 ± 0.1 mm Hg (3/16 in) and 92.0 ± 0.2 mm Hg (1/4 in) at 35°C. Pressure drops across arterial line tubing were main CPB circuit pressure drops. High flow rate, hypothermia, small diameter arterial tubing. and arterial cannula created more hemodynamic energy at the preoxygenator site, but energy loss across CPB circuit also increased. Although small diameter (<1/4 in ID) arterial tubing may decrease total CPB priming volume, it also led to significantly higher circuit pressure, higher pressure drop, and more hemodynamic energy loss across CPB circuit. Larger diameter arterial cannula had less pressure drop and allowed more hemodynamic energy delivery to the patient.
Keywords: Arterial cannula; Arterial tubing; Cardiopulmonary bypass; Neonates; Pediatrics.
Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Similar articles
-
In Vitro Hemodynamic Evaluation of Five 6 Fr and 8 Fr Arterial Cannulae in Simulated Neonatal Cardiopulmonary Bypass Circuits.Artif Organs. 2016 Jan;40(1):56-64. doi: 10.1111/aor.12579. Epub 2015 Nov 2. Artif Organs. 2016. PMID: 26526567
-
Evaluation of neonatal membrane oxygenators with respect to gaseous microemboli capture and transmembrane pressure gradients.Artif Organs. 2010 Nov;34(11):923-9. doi: 10.1111/j.1525-1594.2010.01109.x. Artif Organs. 2010. PMID: 21092035
-
Impact of Pulsatility and Flow Rates on Hemodynamic Energy Transmission in an Adult Extracorporeal Life Support System.Artif Organs. 2015 Jul;39(7):E127-37. doi: 10.1111/aor.12484. Epub 2015 Apr 20. Artif Organs. 2015. PMID: 25894993
-
Pulsatile perfusion during cardiopulmonary bypass procedures in neonates, infants, and small children.ASAIO J. 2007 Nov-Dec;53(6):706-9. doi: 10.1097/MAT.0b013e318158e3f9. ASAIO J. 2007. PMID: 18043152 Review.
-
Quantification of pressure-flow waveforms and selection of components for the pulsatile extracorporeal circuit.J Extra Corpor Technol. 2009 Mar;41(1):P20-5. J Extra Corpor Technol. 2009. PMID: 19361036 Free PMC article. Review.
Cited by
-
Effects of cardiopulmonary bypass with low-priming volume on clinical outcomes in children undergoing congenital heart disease surgery.J Cardiothorac Surg. 2020 May 27;15(1):118. doi: 10.1186/s13019-020-01151-w. J Cardiothorac Surg. 2020. PMID: 32460864 Free PMC article.
-
Effects of albumin and crystalloid priming strategies on red blood cell transfusions in on-pump cardiac surgery: a network meta-analysis.BMC Anesthesiol. 2024 Jan 16;24(1):26. doi: 10.1186/s12871-024-02414-y. BMC Anesthesiol. 2024. PMID: 38229019 Free PMC article.
-
Performance Evaluation of Geometrically Different Pediatric Arterial Cannulae in a Pediatric Cardiopulmonary Bypass Model.Braz J Cardiovasc Surg. 2023 Nov 9;39(1):e20230110. doi: 10.21470/1678-9741-2023-0110. Braz J Cardiovasc Surg. 2023. PMID: 37947184 Free PMC article.
-
Functional Performance of Different Venous Limb Options in Simulated Neonatal/Pediatric Cardiopulmonary Bypass Circuits.Braz J Cardiovasc Surg. 2018 May-Jun;33(3):224-232. doi: 10.21470/1678-9741-2018-0074. Braz J Cardiovasc Surg. 2018. PMID: 30043914 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources