Treatment of an unusual complication of transfemoral TAVI with a new technique: successful occlusion of ventricular septal defect by opening the closure device in the ascending aorta
- PMID: 25633372
- DOI: 10.5830/CVJA-2014-077
Treatment of an unusual complication of transfemoral TAVI with a new technique: successful occlusion of ventricular septal defect by opening the closure device in the ascending aorta
Abstract
Ventricular septal defect (VSD) is a rare complication of transcatheter aortic valve implantation (TAVI) via the transfemoral approach. Aetiological factors leading to VSD have been reported as post-balloon dilatation, oversized prosthesis implantation, and severe calcification of the aorta. However, we present a case of VSD occurring after TAVI with an Edwards Sapien XT prosthesis without any distinct aetiological factors. We used a new technique for closure of the significant VSD; opening the left ventricular disc of the closure device in the ascending aorta and successfully implanting the device without any damage to the bioprosthetic valve.
Similar articles
-
Percutaneous Ventricular Septal Defect Closure After Sapien 3 Transcatheter Aortic Valve Replacement.JACC Cardiovasc Interv. 2015 Jun;8(7):e109-10. doi: 10.1016/j.jcin.2015.01.034. Epub 2015 May 20. JACC Cardiovasc Interv. 2015. PMID: 26003030 No abstract available.
-
Rare complication of ventricular septal defect in three patients following transcatheter aortic valve replacement.Catheter Cardiovasc Interv. 2014 Feb 15;83(3):497-501. doi: 10.1002/ccd.25207. Catheter Cardiovasc Interv. 2014. PMID: 24123754
-
Transcatheter closure of an iatrogenic ventricular septal defect after SAPIEN valve implantation.Catheter Cardiovasc Interv. 2015 Feb 15;85(3):E90-4. doi: 10.1002/ccd.25503. Epub 2014 May 7. Catheter Cardiovasc Interv. 2015. PMID: 24753435
-
Transcatheter Closure of Iatrogenic VSDs after Aortic Valve Replacement Surgery: 2 Case Reports and a Literature Review.Tex Heart Inst J. 2016 Aug 1;43(4):329-33. doi: 10.14503/THIJ-15-5083. eCollection 2016 Aug. Tex Heart Inst J. 2016. PMID: 27547145 Free PMC article. Review.
-
Aortic annular rupture during TAVR: Mini review.Cardiovasc Revasc Med. 2016 Apr-May;17(3):199-201. doi: 10.1016/j.carrev.2016.03.005. Epub 2016 Mar 12. Cardiovasc Revasc Med. 2016. PMID: 27157295 Review. No abstract available.
Cited by
-
Percutaneous closure of an iatrogenic ventricular septal defect following transcatheter aortic valve implantation.J Cardiol Cases. 2017 Aug 12;16(5):165-167. doi: 10.1016/j.jccase.2017.07.004. eCollection 2017 Nov. J Cardiol Cases. 2017. PMID: 30279826 Free PMC article.
-
More in, better out? Successful valve-in-valve procedure of an iatrogenic ventricular septal defect following transcatheter aortic valve replacement: a case report.Eur Heart J Case Rep. 2021 May 12;5(5):ytab097. doi: 10.1093/ehjcr/ytab097. eCollection 2021 May. Eur Heart J Case Rep. 2021. PMID: 34013160 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources