Staging of keratoconus indices regarding tomography, topography, and biomechanical measurements
- PMID: 25634534
- DOI: 10.1016/j.ajo.2015.01.014
Staging of keratoconus indices regarding tomography, topography, and biomechanical measurements
Abstract
Purpose: To derive limits of metric keratoconus indices for classification into keratoconus stages.
Design: Validity and reliability analysis of diagnostic tools.
Methods: A total of 126 patients from the keratoconus center of Homburg/Saar were evaluated with respect to Amsler criteria, using Pentacam (Keratoconus Index [KI], Topographic Keratoconus Classification [TKC]), Topographic Modeling System (Smolek/Klyce, Klyce/Maeda), and Ocular Response Analyzer (Keratoconus Match Probability [KMP], Keratoconus Match Index [KMI]). Mean value, standard deviation, 90% confidence interval, and the Youden J index for definition of the thresholds were evaluated.
Results: For separation of keratoconus stages 0/1/2/3/4 we derived the following optimum thresholds: for KI 1.05/1.15/1.31/1.49 and for KMI 0.77/0.32/-0.08/-0.3. For Smolek/Klyce and Klyce/Maeda high standard deviations and overlapping confidence intervals were found; therefore no discrete thresholds could be defined. Nevertheless, for them we still found a good sensitivity and specificity in discriminating between healthy (stage 0) and keratoconus (stages 2-4) eyes in comparison with the other indices.
Conclusions: We derived thresholds for the metric keratoconus indices KI and KMI, which allow classification of keratoconus stages. These now need to be validated in clinical use. Smolek/Klyce and Klyce/Maeda were not sufficiently sensitive to allow classification into individual stages, but these indices did show a good specificity and sensitivity in discriminating between keratoconus and healthy eyes.
Copyright © 2015 Elsevier Inc. All rights reserved.
Similar articles
-
Biomechanical diagnosis of keratoconus: evaluation of the keratoconus match index and the keratoconus match probability.Acta Ophthalmol. 2013 Jun;91(4):e258-62. doi: 10.1111/aos.12056. Epub 2013 Apr 5. Acta Ophthalmol. 2013. PMID: 23557430
-
Diagnostic capacity of biomechanical indices from a dynamic bidirectional applanation device in pellucid marginal degeneration.J Cataract Refract Surg. 2014 Jun;40(6):1006-12. doi: 10.1016/j.jcrs.2014.03.018. Epub 2014 Apr 26. J Cataract Refract Surg. 2014. PMID: 24774010
-
Diagnostic capacity of the keratoconus match index and keratoconus match probability in subclinical keratoconus.J Cataract Refract Surg. 2014 Jun;40(6):999-1005. doi: 10.1016/j.jcrs.2013.08.064. Epub 2014 Apr 6. J Cataract Refract Surg. 2014. PMID: 24713585
-
[A comparison of device-based diagnostic methods for keratoconus].Klin Monbl Augenheilkd. 2014 Dec;231(12):1215-23. doi: 10.1055/s-0034-1368599. Epub 2014 Jul 15. Klin Monbl Augenheilkd. 2014. PMID: 25025647 German.
-
Can retinoscopy keep up in keratoconus diagnosis?Cont Lens Anterior Eye. 2015 Aug;38(4):234-9. doi: 10.1016/j.clae.2015.01.015. Epub 2015 Feb 25. Cont Lens Anterior Eye. 2015. PMID: 25731818
Cited by
-
Smart decision support system for keratoconus severity staging using corneal curvature and thinnest pachymetry indices.Eye Vis (Lond). 2024 Jul 8;11(1):28. doi: 10.1186/s40662-024-00394-1. Eye Vis (Lond). 2024. PMID: 38978067 Free PMC article.
-
The importance of corneal biomechanics in assessing first degree family members of keratoconus patients.Rom J Ophthalmol. 2018 Apr-Jun;62(2):149-154. Rom J Ophthalmol. 2018. PMID: 30206559 Free PMC article.
-
Complementary Keratoconus Indices Based on Topographical Interpretation of Biomechanical Waveform Parameters: A Supplement to Established Keratoconus Indices.Comput Math Methods Med. 2017;2017:5293573. doi: 10.1155/2017/5293573. Epub 2017 Feb 7. Comput Math Methods Med. 2017. PMID: 28270858 Free PMC article.
-
Corneal Biomechanics in Ectatic Diseases: Refractive Surgery Implications.Open Ophthalmol J. 2017 Jul 31;11:176-193. doi: 10.2174/1874364101711010176. eCollection 2017. Open Ophthalmol J. 2017. PMID: 28932334 Free PMC article.
-
Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study.BMC Ophthalmol. 2023 Jan 26;23(1):36. doi: 10.1186/s12886-023-02767-x. BMC Ophthalmol. 2023. PMID: 36703165 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical