Field testing and comparative evaluation of rapid, visually read screening assays for antibody to human immunodeficiency virus
- PMID: 2564112
- DOI: 10.1016/s0140-6736(89)91610-3
Field testing and comparative evaluation of rapid, visually read screening assays for antibody to human immunodeficiency virus
Abstract
Five rapid, visually read assays for detection of antibody against human immunodeficiency virus (HIV) were evaluated on fresh serum samples from 4000 prospective blood donors at Mama Yemo Hospital, Kinshasa, Zaïre. The sensitivity of the assays, based on 214 specimens positive by western blot, ranged from 84.6% to 99.1%. The specificity, based on 3664 samples negative by enzyme-linked immunosorbent assay (ELISA) or western blot, ranged from 92.7% to 98.8%. Three readers scored each test result independently; disagreement about test interpretation occurred in 1.2-8.3% of the specimens. There was no correlation between assay performance and assay principle (agglutination or dot immunobinding) or antigen source (viral lysate or recombinant). Assays such as these can be readily implemented in a developing country transfusion centre, where blood screening by ELISA is not practicable.
PIP: 5 rapid, visually read blood screening tests for HIV that require no refrigeration or electric power were evaluated on 4000 sera at Mama Yemo Hospital, Kinshasa, Zaire. The tests were 2 1st generation agglutination assays (Quick PHT-HIV, Salck Industries, Sao Paulo, Brazil; and Serodia-HIV, Fujirebio, Tokyo) a 2nd-generation agglutination assay (Recombigen-LA-HIV-1, Cambridge Bio-Science, Worcester, Mass), and 2 2nd generation dot immunobinding assays (HIVCHEK, DuPont, Wilmington, Del, and UCD DOT EIA, University of Ca, Davis, Ca). The blood samples were tested in batches, on east test, confirmed by ELISA, and 1 in 10 by Western blotting. 214 of the sera were positive by Western blot (5.5% prevalence). True positives ranged from 181-212, 211 by ELISA. False positives ranged from 44-267, 71 by ELISA. False positives ranged from 44-267, 71 by ELISA. Variability among test readers was least with the DuPont assay (1.2%), and gretest with the UCD (8.3%). Best agreement between initial and repeat tests was obtained for the DuPont and Fujirebio assays. All 5 tests could be done within 2 hours, the DuPont and Cambridge tests within 10 minutes. None of the tests required electrically powered equipment, although both the Fujirebio and Salck test needed precision pipettes. Technicians found the DuPont test the easiest to read, and the UCD the most difficult. The UCD test seemed to be the most difficult to learn. The DuPont and Fujirebio assays were the most sensitive and specific, and were considered the easiest to perform and interpret. While the DuPont test took only 5 minutes compared to 2 hours for the Fujirebio, the Fujirebio test was the cheapest.
Comment in
-
Sensitivity of HIV screening tests.Lancet. 1989 Jun 10;1(8650):1325-6. doi: 10.1016/s0140-6736(89)92715-3. Lancet. 1989. PMID: 2566849 No abstract available.
Similar articles
-
Evaluation of enzyme-linked immunosorbent and alternative assays for detection of HIV antibodies using panels of Brazilian sera.Rev Inst Med Trop Sao Paulo. 1990 Mar-Apr;32(2):96-100. doi: 10.1590/s0036-46651990000200006. Rev Inst Med Trop Sao Paulo. 1990. PMID: 2095632
-
Field evaluation of alternative HIV testing strategy with a rapid immunobinding assay and an agglutination assay.Lancet. 1991 Jun 1;337(8753):1328-31. doi: 10.1016/0140-6736(91)92991-a. Lancet. 1991. PMID: 1674306
-
Accurate detection of maternal antibodies to HIV in newborn whole blood dried on filter paper.AIDS. 1993 Apr;7(4):483-8. doi: 10.1097/00002030-199304000-00005. AIDS. 1993. PMID: 8507414
-
Laboratory diagnosis of HIV infection in Papua New Guinea.P N G Med J. 1996 Sep;39(3):200-4. P N G Med J. 1996. PMID: 9795563 Review.
-
Anti-HIV antibody testing: procedures and precautions.Am J Infect Control. 1989 Aug;17(4):213-25. doi: 10.1016/0196-6553(89)90131-4. Am J Infect Control. 1989. PMID: 2672903 Review.
Cited by
-
Evaluation of Dot Immunogold Filtration Assay (DIGFA) By Recombinant Protein EPC1 for Anti- Echinococcus granulosus IgG Antibody.Iran J Parasitol. 2015 Jan-Mar;10(1):30-8. Iran J Parasitol. 2015. PMID: 25904943 Free PMC article.
-
Ocular problems in AIDS.Int Ophthalmol. 1990 May;14(3):165-72. doi: 10.1007/BF00158314. Int Ophthalmol. 1990. PMID: 2188917 Review.
-
Comparative evaluation of 36 commercial assays for detecting antibodies to HIV.Bull World Health Organ. 1991;69(6):753-60. Bull World Health Organ. 1991. PMID: 1786625 Free PMC article.
-
Sensitivity of five rapid HIV tests on oral fluid or finger-stick whole blood: a real-time comparison in a healthcare setting.PLoS One. 2010 Jul 19;5(7):e11581. doi: 10.1371/journal.pone.0011581. PLoS One. 2010. PMID: 20657834 Free PMC article.
-
A rapid ultrasound particle agglutination method for HIV antibody detection: Comparison with conventional rapid HIV tests.J Virol Methods. 2017 Nov;249:38-47. doi: 10.1016/j.jviromet.2017.08.015. Epub 2017 Aug 24. J Virol Methods. 2017. PMID: 28843787 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources