Augmented glenoid component designs for type B2 erosions: a computational comparison by volume of bone removal and quality of remaining bone
- PMID: 25648971
- DOI: 10.1016/j.jse.2014.12.018
Augmented glenoid component designs for type B2 erosions: a computational comparison by volume of bone removal and quality of remaining bone
Abstract
Background: The purpose of this computational modeling study was to compare the volume of glenoid bone removal required to implant 3 augmented component designs for management of B2 erosions. In addition, we assessed bone quality of the supporting bone directly beneath the implants by measuring bone density and porosity.
Methods: Three augmented component designs—full-wedge, posterior-wedge, and posterior-step—were studied by virtual implantation in a cohort of 16 patients with B2 glenoids. B2 retroversion was corrected to 0° and 10°. The outcome variables were the volume of glenoid bone removal required for implantation and the density and porosity of the bone immediately beneath the implant.
Results: Implant design had a significant effect on the volume of bone removal (P < .001). When correcting to 0°, the posterior-wedge implant removed less bone than the posterior-step (P < .001) and the full-wedge (P = .004). At 10° retroversion, the posterior-wedge removed less bone (P = .029) than the posterior-step but was no different than the full-wedge (P = .143). The residual glenoid bone density with the posterior-wedge was significantly greater than with the posterior-step (P = .048), with no other significant differences (P > .05). Residual glenoid bone porosity was not significantly different between implants (P > .262).
Conclusions: Augmented components can provide a bone-preserving option for B2 glenoid management. Substantial variations in the volume of bone removal and the quality of the remaining glenoid bone were found between 3 different designs of augmented implants. Simulations with the posterior-wedge implant resulted in substantially less glenoid bone removal, with the remaining supporting bone being of better quality.
Keywords: B2; Biconcave; arthroplasty; augmented implant; osteoarthritis; replacement; shoulder.