A fully biodegradable patent ductus arteriosus occlude
- PMID: 25649512
- DOI: 10.1007/s10856-015-5422-6
A fully biodegradable patent ductus arteriosus occlude
Abstract
The purpose of this study was to develop a fully degradable occluder for the closure of PDA, which can be deployed percutaneously. The blends of biodegradable poly(ε-caprolactone) and poly(L-lactide-co-ε-caprolactone) with various compositions were studied as the potential material. The mechanical properties, i.e. elastic modulus and strain recovery, of the blends could be largely tailored by changing the continuous phase component. Moreover, the suitable blends were selected to fabricate a prototype and its in vitro biodegradation rate and blood compatibility, was evaluated. The current results indicate that no adverse effect on the platelet and leukocyte components of the blood. Biocompatibility implantation studies of the device showed acceptable tissue response. Finally, an artificial PDA conduit was created in a pig, and the device deployment was tested from a sheath: the device recovered within 2-3 min of unsheathing and fully sealed the conduit.
Similar articles
-
The mechanical behavior and biocompatibility of polymer blends for Patent Ductus Arteriosus (PDA) occlusion device.J Mech Behav Biomed Mater. 2014 Aug;36:143-60. doi: 10.1016/j.jmbbm.2014.04.012. Epub 2014 May 2. J Mech Behav Biomed Mater. 2014. PMID: 24846584
-
[Effect of the valved stent on closure of patent ductus arteriosus in a miniswine model].Zhonghua Xin Xue Guan Bing Za Zhi. 2009 May;37(5):458-61. Zhonghua Xin Xue Guan Bing Za Zhi. 2009. PMID: 19781226 Chinese.
-
Feasibility of poly (ϵ-caprolactone-co-DL-lactide) as a biodegradable material for in situ forming implants: evaluation of drug release and in vivo degradation.Drug Dev Ind Pharm. 2015 Feb;41(2):342-52. doi: 10.3109/03639045.2013.866140. Epub 2013 Dec 9. Drug Dev Ind Pharm. 2015. PMID: 24320881
-
Tissue repair in neonatal and paediatric surgery: Analysis of infection in surgical implantation of synthetic resorbable biomaterials.Biomed Mater Eng. 2018;29(6):799-808. doi: 10.3233/BME-181024. Biomed Mater Eng. 2018. PMID: 30282335 Review.
-
Strategy on biological evaluation for biodegradable/absorbable materials and medical devices.Biomed Mater Eng. 2018;29(3):269-278. doi: 10.3233/BME-181735. Biomed Mater Eng. 2018. PMID: 29578468 Review.
Cited by
-
[Establishment of the patent ductus arteriosus model in preterm rats].Zhongguo Dang Dai Er Ke Za Zhi. 2016 Apr;18(4):372-5. doi: 10.7499/j.issn.1008-8830.2016.04.018. Zhongguo Dang Dai Er Ke Za Zhi. 2016. PMID: 27097586 Free PMC article. Chinese.
-
Translation in cardiovascular stents and occluders: From biostable to fully degradable.Bioeng Transl Med. 2017 Jul 19;2(2):156-169. doi: 10.1002/btm2.10066. eCollection 2017 Jun. Bioeng Transl Med. 2017. PMID: 29313029 Free PMC article. Review.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources