In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters
- PMID: 25657531
- PMCID: PMC4313483
- DOI: 10.4103/0972-0707.148899
In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters
Abstract
Background: There is a growing interest about electronic apex locators for working length determination. There are several studies dealing with their performance in different conditions.
Aims: The aim of this study is to evaluate the accuracy of Root ZX and Raypex 6 in teeth with different apical diameters.
Materials and methods: Actual working length (AWL) of 80 single rooted teeth were determined as 0.5 mm short of apical foramen. The teeth were divided into 4 groups (n = 20). First group (G 0) included teeth with mature apices. Root canals of the other groups (G 32, G 57 and G 72) were enlarged until apical sizes of 0.32, 0.57 and 0.72 mm were obtained. Samples were embedded in alginate and electronic measurements (EM) were performed.
Statistical analysis: was achieved with Fisher exact test.
Results: Both devices revealed a high rate of success in G 0 and G 32. Their accuracy decreased significantly in G 57 and G 72 groups (P < 0.05). Intra-group results of Root ZX and Raypex 6 were similar (P > 0.05).
Conclusions: Root ZX and Raypex 6 are reliable in teeth with mature apices. At foramen diameters exceeding 0.57 mm, their accuracy is susceptible.
Keywords: Apical constriction; Raypex 6; Root ZX; electronic apex locator; working length determination.
Conflict of interest statement
Similar articles
-
Accuracy of Four Apex Locators in Premolars with Root Resorption: An In vitro Study.Iran Endod J. 2022 Winter;17(1):20-26. doi: 10.22037/iej.v17i1.35498. Iran Endod J. 2022. PMID: 36703871 Free PMC article.
-
Impact of Apical Patency on Accuracy of Electronic Apex Locators: In Vitro Study.J Endod. 2020 Apr;46(4):509-514. doi: 10.1016/j.joen.2020.01.010. Epub 2020 Feb 20. J Endod. 2020. PMID: 32089336
-
Accuracy of Different Apex Locators in Teeth with Simulated Apical Root Resorption: an In Vitro Study.Folia Med (Plovdiv). 2018 Dec 1;60(4):624-631. doi: 10.2478/folmed-2018-0033. Folia Med (Plovdiv). 2018. PMID: 31188764
-
Determination of working length for teeth with wide or immature apices: a review.Int Endod J. 2013 Jun;46(6):483-91. doi: 10.1111/iej.12032. Epub 2012 Dec 7. Int Endod J. 2013. PMID: 23216284 Review.
-
Comparative Evaluation of Electronic Apex Locators and Radiovisiography for Working Length Determination in Primary Teeth in vivo.Int J Clin Pediatr Dent. 2016 Apr-Jun;9(2):118-23. doi: 10.5005/jp-journals-10005-1346. Epub 2016 Jun 15. Int J Clin Pediatr Dent. 2016. PMID: 27365931 Free PMC article. Review.
Cited by
-
Accuracy of Four Apex Locators in Premolars with Root Resorption: An In vitro Study.Iran Endod J. 2022 Winter;17(1):20-26. doi: 10.22037/iej.v17i1.35498. Iran Endod J. 2022. PMID: 36703871 Free PMC article.
-
Accuracy of Root ZX Electronic Apex Locator in Relation to Two Different Employment Protocols: An In Vitro Study.Cureus. 2023 Sep 4;15(9):e44659. doi: 10.7759/cureus.44659. eCollection 2023 Sep. Cureus. 2023. PMID: 37671079 Free PMC article.
-
Factors influencing the accuracy of electronic apex locators: A scoping review.Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025. Dent Res J (Isfahan). 2025. PMID: 40655901 Free PMC article. Review.
-
Comparative evaluation of the accuracy of two electronic apex locators in detecting simulated incomplete vertical root fractures: An in vitro stereomicroscopic study.J Conserv Dent Endod. 2024 May;27(5):540-544. doi: 10.4103/JCDE.JCDE_132_24. Epub 2024 May 10. J Conserv Dent Endod. 2024. PMID: 38939546 Free PMC article.
-
A comparative clinical evaluation of accuracy of six apex locators with intraoral periapical radiograph in multirooted teeth: An in vivo study.J Conserv Dent. 2017 Jul-Aug;20(4):264-268. doi: 10.4103/JCD.JCD_131_17. J Conserv Dent. 2017. PMID: 29259365 Free PMC article.
References
-
- Heo MS, Han DH, An BM, Huh KH, Yi WJ, Lee SS, et al. Effect of ambient light and bit depth of digital radiograph on observer performance in determination of endodontic file positioning. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:239–44. - PubMed
-
- Stober EK, Duran-Sindreu F, Mercade M, Vera J, Bueno R, Roig M. An evaluation of root ZX and iPex apex locators: An in vivo study. J Endod. 2011;37:608–10. - PubMed
-
- Elayouti A, Weiger R, Lost C. Frequency of overinstrumentation with an acceptable radiographic working length. J Endod. 2001;27:49–52. - PubMed
-
- Mancini M, Felici R, Conte G, Costantini M, Cianconi L. Accuracy of three electronic apex locators in anterior and posterior teeth: An ex vivo study. J Endod. 2011;37:684–7. - PubMed
-
- Tinaz AC, Sevimli LS, Gorgul G, Turkoz EG. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod. 2002;28:160–2. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous