In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters

Affiliations

In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters

Ugur Aydin et al. J Conserv Dent. 2015 Jan-Feb.

Abstract

Background: There is a growing interest about electronic apex locators for working length determination. There are several studies dealing with their performance in different conditions.

Aims: The aim of this study is to evaluate the accuracy of Root ZX and Raypex 6 in teeth with different apical diameters.

Materials and methods: Actual working length (AWL) of 80 single rooted teeth were determined as 0.5 mm short of apical foramen. The teeth were divided into 4 groups (n = 20). First group (G 0) included teeth with mature apices. Root canals of the other groups (G 32, G 57 and G 72) were enlarged until apical sizes of 0.32, 0.57 and 0.72 mm were obtained. Samples were embedded in alginate and electronic measurements (EM) were performed.

Statistical analysis: was achieved with Fisher exact test.

Results: Both devices revealed a high rate of success in G 0 and G 32. Their accuracy decreased significantly in G 57 and G 72 groups (P < 0.05). Intra-group results of Root ZX and Raypex 6 were similar (P > 0.05).

Conclusions: Root ZX and Raypex 6 are reliable in teeth with mature apices. At foramen diameters exceeding 0.57 mm, their accuracy is susceptible.

Keywords: Apical constriction; Raypex 6; Root ZX; electronic apex locator; working length determination.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: None declared.

Similar articles

Cited by

References

    1. Heo MS, Han DH, An BM, Huh KH, Yi WJ, Lee SS, et al. Effect of ambient light and bit depth of digital radiograph on observer performance in determination of endodontic file positioning. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:239–44. - PubMed
    1. Stober EK, Duran-Sindreu F, Mercade M, Vera J, Bueno R, Roig M. An evaluation of root ZX and iPex apex locators: An in vivo study. J Endod. 2011;37:608–10. - PubMed
    1. Elayouti A, Weiger R, Lost C. Frequency of overinstrumentation with an acceptable radiographic working length. J Endod. 2001;27:49–52. - PubMed
    1. Mancini M, Felici R, Conte G, Costantini M, Cianconi L. Accuracy of three electronic apex locators in anterior and posterior teeth: An ex vivo study. J Endod. 2011;37:684–7. - PubMed
    1. Tinaz AC, Sevimli LS, Gorgul G, Turkoz EG. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod. 2002;28:160–2. - PubMed