Objective: To investigate a potential relationship between implantable cardioverter defibrillator (ICD) therapies and daily geomagnetic activity (GMA) recorded in a large database.
Patients and methods: The ALTITUDE database, derived from the Boston Scientific LATITUDE remote monitoring system, was retrospectively analyzed for the frequency of ICD therapies. Daily GMA was expressed as the planetary K-index and the integrated A-index and was graded as levels I (quiet), II (unsettled), III (active), and IV (storm).
Results: A daily mean ± SD of 59,468±11,397 patients were monitored between January 1, 2009, and May 15, 2012. The distribution of days according to GMA was as follows: level I, 924/1231 (75%); level II, 226/1231 (18%); level III, 60/1231 (5%); and level IV, 21/1231 (2%). The daily mean ± SD numbers of ICD shocks received per 1000 active patients in the database were 1.29±0.47, 1.17±0.46, 1.03±0.37, and 0.94±0.29 on level I, II, III, and IV days, respectively; the daily mean ± SD sums of shocks and antitachycardia pacing therapies were 9.29±2.86, 8.46±2.45, 7.92±1.80, and 7.83±2.28 on quiet, unsettled, active, and storm days, respectively. A significant inverse relationship between GMA and frequency of ICD therapies was identified, with the most pronounced difference between level I and level IV days (P<.001 for shocks; P=.008 for shocks + antitachycardia pacing).
Conclusion: In a large-scale cohort analysis, ICD therapies were delivered less frequently on days of higher GMA, confirming the previous pilot data and suggesting that higher GMA does not pose an increased risk of arrhythmias using ICD therapies as a surrogate marker. Further studies are needed to gain an in-depth understanding of the underlying mechanisms.
Copyright © 2015 Mayo Foundation for Medical Education and Research. Published by Elsevier Inc. All rights reserved.
Conflict of interest statement
All authors have completed and submitted the ICMJE Form for Disclosure of Potential Conflicts of Interest.
Figures
References
-
- Lampert R, Joska T, Burg MM, Batsford WP, McPherson CA, Jain D. Emotional and physical precipitants of ventricular arrhythmia. Circulation. 2002;106:1800–1805. - PubMed
-
- Barnes BJ, Hollands JM. Drug-induced arrhythmias. Critical care medicine. 2010;38:S188–S197. - PubMed
-
- Rozelot JP. Lect. Notes Phys. Vol. 699. Berlin Heidelberg: Springer; 2006. Solar and Heliospheric Origins of Space Weather Phenomena.
-
- Stoupel E. Sudden cardiac deaths and ventricular extrasystoles on days with four levels of geomagnetic activity. Journal of basic and clinical physiology and pharmacology. 1993;4:357–366. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical