Comparative Study
. 2015 Mar;42(3):120-2.
doi: 10.1097/OLQ.0000000000000240.

Confirmation of high specificity of an automated enzyme immunoassay test for serological diagnosis of syphilis: retrospective evaluation versus results after implementation

Affiliations
Comparative Study

Confirmation of high specificity of an automated enzyme immunoassay test for serological diagnosis of syphilis: retrospective evaluation versus results after implementation

Laura van Dommelen et al. Sex Transm Dis. 2015 Mar.

Abstract

Background: The optimal algorithm for serological syphilis screening is still a matter of debate. We have previously evaluated the performance of the Bioelisa Syphilis 3.0, using a selection of archived sera, and in this study compare these results with the Bioelisa results after clinical implementation.

Methods: All Bioelisa Syphilis 3.0 results obtained since clinical implementation were analyzed. Bioelisa-positive or borderline samples were retested using Treponema pallidum particle agglutination, rapid plasma reagin test, fluorescent treponemal antibody-absorption test, and/or immunoblot. On sera sent in together with cerebrospinal fluid, occasionally both the T. pallidum particle agglutination and Bioelisa were performed.

Results: The Bioelisa was performed on 14,622 sera. Bioelisa-positive samples, which were not retested by the previously described assays, were withdrawn from the database (n = 36). In 1.3% of the samples (187/14,586), the Bioelisa was positive or borderline and, ultimately, 115 sera were considered true positive (prevalence 0.8%). The specificity of the Bioelisa was 99.5%.

Conclusions: Based on the results of all performed diagnostic assays, the specificity of the Bioelisa of 99.5% is very consistent with that found in the initial study (100%; 95% confidence interval was 98.0%-100%). Interpreting (positive) test results is difficult in the absence of a gold standard, especially when the disease prevalence is low. Results should be viewed in the light of the patients' characteristics.