Preconcentration of lead using solidification of floating organic drop and its determination by electrothermal atomic absorption spectrometry
- PMID: 25685441
- PMCID: PMC4293875
- DOI: 10.1016/j.jare.2012.07.002
Preconcentration of lead using solidification of floating organic drop and its determination by electrothermal atomic absorption spectrometry
Abstract
A simple microextraction method based on solidification of a floating organic drop (SFOD) was developed for preconcentration of lead prior to its determination by electrothermal atomic absorption spectrometry (ETAAS). Ammonium pyrolidinedithiocarbamate (APDC) was used as complexing agent, and the formed complex was extracted into a 20 μL of 1-undecanol. The extracted complex was diluted with ethanol and injected into a graphite furnace. An orthogonal array design (OAD) with OA16 (4(5)) matrix was employed to study the effects of different parameters such as pH, APDC concentration, stirring rate, sample solution temperature and the exposure time on the extraction efficiency. Under the optimized experimental conditions the limit of detection (based on 3 s) and the enhancement factor were 0.058 μg L(-1) and 113, respectively. The relative standard deviation (RSD) for 8 replicate determinations of 1 μg L(-1) of Pb was 8.8%. The developed method was validated by the analysis of certified reference materials and was successfully applied to the determination of lead in water and infant formula base powder samples.
Keywords: Ammonium pyrolidinedithiocarbamate; Electrothermal atomic absorption; Lead; Microextraction; Milk powder sample; Orthogonal array design.
Similar articles
-
Ultra-trace determination of lead in water and food samples by using ionic liquid-based single drop microextraction-electrothermal atomic absorption spectrometry.Anal Chim Acta. 2009 Jun 30;644(1-2):48-52. doi: 10.1016/j.aca.2009.04.029. Epub 2009 May 3. Anal Chim Acta. 2009. PMID: 19463561
-
Preconcentration and speciation of arsenic in water specimens by the combination of solidification of floating drop microextraction and electrothermal atomic absorption spectrometry.Talanta. 2010 Apr 15;81(1-2):197-201. doi: 10.1016/j.talanta.2009.11.056. Epub 2009 Nov 29. Talanta. 2010. PMID: 20188908
-
Ionic liquid-based, single-drop microextraction for preconcentration of cobalt before its determination by electrothermal atomic absorption spectrometry.J AOAC Int. 2010 May-Jun;93(3):985-91. J AOAC Int. 2010. PMID: 20629404
-
A novel separation/preconcentration system based on solidification of floating organic drop microextraction for determination of lead by graphite furnace atomic absorption spectrometry.Anal Chim Acta. 2008 Aug 15;623(2):163-7. doi: 10.1016/j.aca.2008.06.033. Epub 2008 Jun 27. Anal Chim Acta. 2008. PMID: 18620920
-
Magnetic materials as sorbents for metal/metalloid preconcentration and/or separation. A review.Anal Chim Acta. 2013 Jul 30;789:1-16. doi: 10.1016/j.aca.2013.04.021. Epub 2013 Apr 29. Anal Chim Acta. 2013. PMID: 23856225 Review.
Cited by
-
Solidified Floating Organic Drop Microextraction for the Detection of Trace Amount of Lead in Various Samples by Electrothermal Atomic Absorption Spectrometry.J Anal Methods Chem. 2017;2017:6268975. doi: 10.1155/2017/6268975. Epub 2017 Jul 20. J Anal Methods Chem. 2017. PMID: 28808599 Free PMC article.
References
-
- Xie Y., Chiba M., Shinohara A., Watanabe H., Inaba Y. Studies on lead-binding protein and interaction between lead and selenium in the human erythrocytes. Ind Health. 1998;36:234–237. - PubMed
-
- Saryan L.A., Zenz C. Lead and its compounds. In: Zenz O.C., Dickerson B., Horvath E.P., editors. Occupational medicine. Mosby-Year Book, Inc.; Missouri: 1994. pp. 506–510.
-
- Huang X., Sillanpa M., Duo B., Gjessing E.T. Water quality in the tibetan plateau: metal contents of four selected rivers. Environ Pollut. 2008;156:270–277. - PubMed
-
- ATSDR, Agency for Toxic Substances and Disease Registry. Toxicological profile for lead, US department of health and human services. Atlanta: Public Health Service; 1999. p. 465.
-
- Dos Santos W.L., dos Santos C.M.M., Costa J.L.O., Andrade H.M.C., Ferreira S.L.C. Multivariate optimization and validation studies in on-line pre-concentration system for lead determination in drinking water and saline waste from oil refinery. Microchem J. 2004;77:123–129.
LinkOut - more resources
Full Text Sources
Other Literature Sources